4.30 PC-300–PC-550 Procedural/Fixed Height/Trauma/Surgical/ OB/GYN/Electric Stretcher Caster
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 96
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
•
If patient security straps are available, use them to secure the base
shroud to the upper frame.
•
If the patient security straps are not available, you can use rope, or
you can remove the base shroud (refer to procedure 4.25).
2. Put the stretcher in the steer position.
3. Using the ratchet and 1/8" hex head bit, remove the setscrew (C) from the
rocker link (E) (see figure 4-51 on page 4-96).
Figure 4-51. Caster Replacement
4. Put the stretcher in the neutral position.
5. Using the hammer and the 5/16" pin punch, gently tap the hex rod (F) out
of the old caster.
6. Using the ratchet and 7/16" socket, remove the caster screw (D).
NOTE:
Use the 13 mm socket to remove the caster screw from the
surgical stretcher
only
.