background image

Description 

 

Ohm Metrics 

Surface Resistance Meter:

 

SRM 500K 

 
The SRM500K is a digital surface resistance test kit designed 
to test all conductive, anti-static and static dissipative sur-
faces for electrical resistivity/resistance according to EOS/
ESD, CECC, ANSI, ASTM test procedures. It accurately and 
quickly measures resistance between two points (RTT) and 
surface to ground resistance (RTG). Suitable for factory au-
dits and test lab evaluations, the Ohm Metrics SRM500K also 
measures ambient temperature and humidity. 
Features include:  

 

10v and 100v range voltage selection

 

 

Automatic zeroing and power shut off

 

 

Two 5 lb disk probes  

 

Built in resistivity probes 

 

Carrying case 

 

The

 

SRM500k meets periodic test requirements per Compli-

ance Verification ESD TR53 and conforms to ANSI/EOS/ESD 
(S4.1, S7.1, S12.1, S2.1). 

 

Summary of Contents for Ohm Metrics SRM500K

Page 1: ...Instruction Manual Surface Resistance Testing Kit Model SRM500K Ohm Metrics...

Page 2: ...1 Ohm Metrics 2 2 Operation Power Option 2 Set Up 2 Parallel Probe Resistivity Testing 3 Probe Resistivity Testing 4 Resistance Between Two Points RTT 4 Surface to Ground Resistance RTG 5 3 Specificat...

Page 3: ...uickly measures resistance between two points RTT and surface to ground resistance RTG Suitable for factory au dits and test lab evaluations the Ohm Metrics SRM500K also measures ambient temperature a...

Page 4: ...ce Verification TR53 All Ohm Metrics products can be calibrated All Ohm Metrics test and measurement products are warranted for 1 full year Power Options The SRM500K can be powered by a 9 volt battery...

Page 5: ...ee of contaminants 1 Place the meter on the surface to be tested 2 Move the switch to desired test voltage position either 10 or 100 volts 3 Press and hold the test button After approximately 15 secon...

Page 6: ...in dependent of a ground point 1 Connect the plug ends of the test leads into the 3 5mm jacks of the meter 2 Connect the banana plugs of the test coil cords into the ends of the 5 pound probes 3 Place...

Page 7: ...to a known ground point 3 Position the probe on the surface to be tested in accor dance with the desired test procedure 4 Press the test button until the resistivity humidity and temperature test val...

Page 8: ...matic ranging Power Supply 9V Battery PP3 120 volt AC Adapter Unit automatically powers off when not in use Probes Two 5 pound disk probes Two 3 parallel surface resistivity probes Read Out LCD alpha...

Page 9: ...7 Specifications Continued Accuracy for 10 volt scale 10e3 10e8 10 RH 90 10e8 10e12 20 RH 90 Accuracy for 100 volt scale 10e3 10e8 10 RH 90 10e9 10e12 20 RH 60...

Page 10: ...the date of shipment Liability is limited to servicing after evalu ating repairing or replacing any product returned to Transforming Technologies The company does not war rant damage due to misuse neg...

Page 11: ...ect their prod ucts and processes from the many serious problems as sociated with static electricity Transforming Technologies offers a wide range of unique and outstanding products to detect protect...

Page 12: ...3719 King Rd Toledo OH 43617 P 419 841 9552 F 419 841 3241 info transforming technologies com...