BFN 802 Performance
Tested in accord with ANSI/ESD STM3.1-2006
Discharge time
in seconds (1000
volts to 100volts)
determined per
EOS/ESD Associa-
tion Standard
NO.3; unit at 18”
above test plate,
fans at high
speed. Discharge
times are slightly
longer for 230VAC,
50Hz unit.
Operating Guide
1. Power Connection
Properly locate and install the ionizer into the proper posi-
tion and connect the power cord. Rotate the on/off, fan
speed knob clockwise to turn on the unit. Continue rotat-
ing the knob to achieve the desired fan speed.
Note:
Do not insert any object into the air inlet or outlet.
Note:
Unit must be properly grounded before operating.
2. Periodic Maintenance
The only maintenance required for BFN series ionizers is
the periodic cleaning of the emitter points, inlet and outlet
finger guards and the fan assembly. Emitter point cleaning
affects the static decay ability of the ionizer and is impor-
tant for maintaining its optimal performance.