                                                              Red LED 

                                                                   

 

 

 

 

 

In normal weigh mode, the LED is flickering: the flicker frequency is 

almost 2Hz in zero, the flicker frequency is almost 1Hz in other 

weight; in setting mode, the LED lighten or black out. 

 

4.3 TI-1500VC Calibration 

 

Note: Every TI-1500VC has Factory Reset (F21) when it out of 

the factory. The 4-20mA and 0-10V has been adjusted, so the 

user shouldn’t repeat F21, otherwise the analog output should 

been adjusted again. 

 

4.3.1 Calibration 

Example: The system result: 3000 x 0.1kg=300kg 

 

(1) Graduations: Enter setting mode, the display show [F1].  

Press “

” key enter, the display show [xxxxx], if now the value 

is [3000], press “

”, back to the main menu. Otherwise, press 

” or “

” key, until the display show [3000], press “

SET

” to 

save the change. Then press “

”, back to the main menu, the 

display show [F1] again. 

(2) Display Divisions: Press “

” key, until the display show [F9].  

Press “

” key enter, the display show [div x], if now the value is 

[div 1], press “

”, back to the main menu. Otherwise, press 

” or “

” key, until the display show [div 1], press “

SET

” to   

-11- 

save the change. Then press “

”, back to the main menu, the 

display show [F9] again. 

(3) Decimal Point: Press “

” key, until the display show [F10].  

Press “

” key enter, the display show [xxx], if now the value is 

[0.0], press “

”, back to the main menu. Otherwise, press “

” 

or “

” key, until the display show [0.0], press “

SET

” to save the 

change. Then press “

”, back to the main menu, the display 

show [F10] again. 

(4) Zero Calibration: Press “

” key, until the display show [F16]. 

Make sure there are no test weights on the platform, press “

” 

key enter, the display show the present zero A/D count. When 

the stabile annunciator turns on, press “

” key to set zero point, 

then press “

SET

” key to save the zero value. 

(5)  Span Calibration: Press “

” key, until the display show [F17]. 

Place the full capacity test weight on the platform, press “

” 

key enter, the display show flash [    0], press “

” key 3 times, 

the display show [    0000], press “

” input [3], now the display 

show [3000], it’s equal to the test weight. When the stabile 

annunciator turns on, press “

SET

” key to save the span value. 

 

Now finishing the calibration, turn off the system. Then turn it on 

again, the system back to the normal weigh mode. 

 

4.3.2 Adjust the Zero Track Band (F3) 

 

Zero Track Band means in stabile state the system will 

automatically zero within the setting zero track range. The system’s 

default zero track value is 1d. If the system use in large wind 

environment or there will be some residues leave on the platform 

every weight, the zero track parameter should be 

-12- 

Summary of Contents for TI-1500VC

Page 1: ...5 Deerfield Parkway Buffalo Grove IL U S A Zip 60089 Tel 01 800 503 9180 01 847 419 9180 Fax 01 847 419 1515 Web Site www transcell net TI 1500VC Weighing Transmitter Setup Operation Manual Ver 3 5C 2018 05 Transcell Technology Inc R ...

Page 2: ... 4 2 Display and Annunciator 10 4 3 TI 1500VC Calibration 11 4 3 1 Calibration 11 4 3 2 Adjust the Zero Track Band F3 12 4 3 3 Adjust the Manual Zero Range F4 13 4 3 4 Adjust the Auto Zero Range F15 13 4 3 5 Adjust DA F25 13 5 RS485 Communication Port 15 5 1 Connecting the RS485 output 15 5 2 Communication Mode of RS485 15 5 3 Data Format TI 1500VC Vision V3 5C 16 5 4 Command Format 17 Hardware Vi...

Page 3: ...uld choose one output channel to operate c Communication port RS485 3 Weighing Resolution III Except 4 20mA 0 10V output 4 Display Keyboard 7 segment LCD 1 LED annunciator 5 keyboards 5 Function Auto zero track Digital Filter Digital Calibration Keep Peak Value 6 Sampling Rate 100Hz 7 Rated Voltage DC 15V 10 500mA 8 Power 4W 9 Operating Temperature 10 55 10 Enclosure Stainless steel water proof 1 ...

Page 4: ...4 should be in short circuit RS 485 terminal resistance is enable 3 3 TI 1500VC Setup Menu 3 1 System menu F menu The section provides more detailed descriptions of the selections found in the system menu chart F menu Factory set F21 defaults are shown in bold with a checkmark F menu Name Code Description Code Value F1 Graduation Specifies number of full scale graduations Value should be consisten...

Page 5: ...normal operation 1 lb 2 kg F9 Display Divisions Determines the desired weigh increments Value should be consistent with legal requirements div 1 div 2 div 5 5 F10 Decimal Pt Determines location of the decimal point 0 0 0 0 00 0 000 0 0000 00 F14 Disable the lb kg key Allows the lb kg key to be disabled so that an operator cannot accidentally press the key and change the displayed units Cr DIS Cr E...

Page 6: ...ence 7 3 2 User menu Changing communication mode and formula is in this menu A menu Name Code Description Code Value A1 Baud Rate Selects the baud rate for data transmission through the serial port 1200 2400 4800 9600 19200 A2 Data Bits and Parity Selects the number of data bits and parity of serial transmission 8n 7O 7E 7n A3 Mode of Serial Transmission Selects when data will be sent out of the s...

Page 7: ...play zero SW3 SET Save the setting value 9 SW4 Display the next sub menu or move left SW5 Display the next sub menu or move right PRINT Print information Enter HOLD Mode 1 Press and Hold SET key while system power on until the display show F1 enter setting mode Turn off the system then turn it on again the system back to normal weigh mode 2 In F or A menu mode press and hold key back to normal wei...

Page 8: ... show F10 Press key enter the display show xxx if now the value is 0 0 press back to the main menu Otherwise press or key until the display show 0 0 press SET to save the change Then press back to the main menu the display show F10 again 4 Zero Calibration Press key until the display show F16 Make sure there are no test weights on the platform press key enter the display show the present zero A D ...

Page 9: ...ameter should not be adjusted again generally If necessary please follow the step 1 Enter F23 to select the type of analog output 2 Install the test circuit Command connect to the device directly 3 Enter F25 press the display show the present DA value 0V or 4mA xxxxx Adjust way key Rough adjust Press once the value increase 50 key Rough adjust Press once the value decrease 50 SET key Save the valu...

Page 10: ... I O port state 3 bits I O port state information Weight Data 8 bits Symbol 1 bit Decimal Point 1 bit xxxxxx 6 bits integer type Weight Units 1 bit K kg L lb State 1 bit 0 Overload M Motion H Hold S Stability Parity bit 1 bit The low byte of the sum of Device address I O port state Weight Data Weight Units and State End symbol 2 bits CR LF Note Weight Data the definition of the decimal point 0 No ...

Page 11: ...VC factory series No CC Transmit Units CH Enter Exit Hold Mode CZ set the system to display zero 17 Hardware Vision Explain Show as follow figure There is a hardware s mark label paste near the power terminal So the different PCB board which marks different sign can not be changed Otherwise the TI 1500VC will be destroyed If the transmitter out of operate because of the hardware error please take ...

Reviews: