) REVIEW OF ELECTRODE STATUS 

This  setting  allows  you  to  review  the  last calibration  information  about 
the electrode conditions. 

1.  There are 5 calibration points for review in the follow order: 

 

SAVED DATA 

NIST Buffer range 

ISO/ Custom Buffer range 

P3.1 

% of Calibration slope 

0.00 to 4.01 

0.00 to 4.50 

P3.2 

% of Calibration slope 

4.02 to 6.86 

4.51 to 7.00 

P3.3 

% of Calibration slope 

6.87 to 9.18 

7.01 to 9.50 

P3.4 

% of Calibration slope 

9.18 to 14.00 

9.51 to 14.00 

P3.5 

Calibration offset value in mili-volt (mV) 

 

2.  Press 

   

 and  display  will  show  value  in  percentage  (%)  for 

P3

P3

P3

P3

.

1

1

1

1

.  Repeat for the next value, in the above table order. 

3.  A  good  electrode  should  have  a  value  of  above  75%  and  below 

115%.  Otherwise, the electrode should be replaced. 

4.  At P3.5, display will show the calibrated offset value in mV.  A good 

electrode should have a value not greater than ±60mV.  Otherwise, 
the electrode should be replaced. 

5.  Press 

   

 to return to 

ELE

ELE

ELE

ELE

 (

P3

P3

P3

P3

.

0

0

0

0

)

 

6.  Press    

 to go to next parameter. 

 

Buf 

(

P4

P4

P4

P4

.

0

0

0

0

) BUFFER CALIBRATION STANDARD 

This  setting  allows  you  to  select  ISO,  NIST  or  customized  buffer 
standard.  Factory preset is ISO standard. 

ISO

  (

i

i

i

iSo

So

So

So

)  buffer  standards  are  pH1.68,  7.00,  4.01,  10.01,  12.45  and 

NIST

  (

n

n

n

ni

i

i

iSt

St

St

St

)  buffer  standards  are  pH1.68,  4.01,  6.86,  9.18  and  12.45, 

which  are  preset  in  this  meter,  and  will  be  automatically  recognized 
during calibration. 
Select  customize  (

CUSt

CUSt

CUSt

CUSt

)  if  you  have  other  standards  other  than  Nist 

buffer solutions.  This will allow you to set each calibration point to the 
standard buffer solution available, which are about to calibrate. 
 
1. Press    

and display will show 

CUSt

CUSt

CUSt

CUSt

 

(

P4

P4

P4

P4

.

1

1

1

1

)

 flashing. 

2. Press    

 once to select 

I

I

I

ISO

SO

SO

SO

,

 n

 n

 n

 ni

i

i

iSt

St

St

St

 or  

CUSt

CUSt

CUSt

CUSt

.  Press     

 and 

display will return to 

BuF

BuF

BuF

BuF

 (

P4

P4

P4

P4

.

0

0

0

0

)

3. Press    

 to go to next parameter. 

Summary of Contents for BP3001

Page 1: ...0 15 0 15 0 1 14 1 14 13 2 13 2 3 12 3 12 11 4 11 4 5 10 5 10 9 6 9 6 7 8 7 8 Operation Manual Professional Benchtop pH meter BP3001 ...

Page 2: ...ATURES Multiple display of pH temperature simultaneously Automatic buffer recognition with built in ISO and NIST standard Manual setting of customize buffer standard 5 points calibration Hold or freeze display function 99 memory with real time clock recording Maximum and minimum data review RS232 online data logging to PC 2 CONTENT Carefully unpack the box It should contain the following items a M...

Page 3: ...Operating humidity Up to 95 RH Meter size 217 x 168 x 58mm Package size Package weight 2 1kg 4 PRODUCT ASSEMBLY Turn anti clockwise to loosen arm to adjust angle turn clockwise to lock arm into position Place meter s left corner onto the STAND BASE to set into position Electrode stand can be set on the right corner of meter as desired Complete the stand assembly by pushing the arm into the top hol...

Page 4: ...ulb is half of the measuring element and must be kept wet to enhance speed of measurement If the sensor is dried out it will take a longer time to get hydrated The other measuring element is the reference element which contain concentrated chemical with a liquid bridge of porous fiber allowing controlled contact with the measuring liquid In order to keep both sensors in ideal condition the electro...

Page 5: ...is being saved MAX Maximum reading of all saved data displayed MIN Minimum reading of all saved data displayed REC Flashing while recall mode is initiated HOLD Display put on hold READY When the READY feature is enabled during measurement it will flash indicating measurement in progress When stopped indicates reading has stabilized 7 5 4 2 1 6 3 Socket for RS232 cable for PC data transfer AC DC Ad...

Page 6: ...CALL mode MODE KEY MOVE UP SELECT KEY 1 Press to switch display mode between pH and mV 2 During setting press to move digit up or next selection MEMORY SAVE KEY MOVE DOWN SELECT KEY 1 Press to save current reading into memory 2 During setting press to move digit down or previous selection CONFIRM KEY MAXIMUM MINIMUM DISPLAY KEY TEMPERATURE SET KEY 1 During CALIBRATION or SETUP setting press to con...

Page 7: ...he PC 4 Under Available ports select the com port meter is connected to 5 Click to select file location and type file name then click save 6 Set Baud rate to 9600 Data bits to 8 Parity to non Stop bits to 1 Flow control to non and Click 7 Press on the meter and display will show out out out out P1 P1 P1 P1 1 1 1 1 flashing PC will indicate transmission in progress 8 When transmission is completed ...

Page 8: ...lue not greater than 60mV Otherwise the electrode should be replaced 5 Press to return to ELE ELE ELE ELE P3 P3 P3 P3 0 0 0 0 6 Press to go to next parameter Buf P4 P4 P4 P4 0 0 0 0 BUFFER CALIBRATION STANDARD This setting allows you to select ISO NIST or customized buffer standard Factory preset is ISO standard ISO i i i i So So So So buffer standards are pH1 68 7 00 4 01 10 01 12 45 and NIST n n...

Page 9: ...ment for temperature Factory preset is in C 1 Press and display will show Unit Unit Unit Unit P7 P7 P7 P7 1 1 1 1 with C C C C flashing 2 Press to change between F F F For C C C C 3 Press to confirm setting display will return to Unit Unit Unit Unit P7 P7 P7 P7 0 0 0 0 4 Press to go to next parameter Rtc P8 P8 P8 P8 0 0 0 0 REAL TIME CLOCK This setting allow you to change the date in Year Month Da...

Page 10: ...de in the acidic range select 2 points between pH0 to 7 If measurements are made in the alkaline range then select between pH7 to 14 If full range is required calibrate at least 3 to 5 calibration points between pH0 to 14 10 2 Buffer standard must be set during SETUP in page 7 BuF BuF BuF BuF P4 P4 P4 P4 0 0 0 0 10 3 Standard solutions are in 3 groups of standards ISO Nist and custom standards ISO...

Page 11: ...01 10 01 12 45 30 1 69 6 98 4 02 9 97 12 29 35 1 69 6 98 4 02 9 93 12 13 40 1 70 6 97 4 03 9 89 11 99 45 1 70 6 97 4 04 9 86 11 84 50 1 71 6 97 4 06 9 83 11 70 10 10 If the sub display does not automatically recognize a buffer it means the electrode could be defective or the buffer solution is wrong Calibration cannot proceed until they are rectified 10 11 When the READY sign appear on the top lef...

Page 12: ...3 25 1 68 4 01 6 86 9 18 12 45 30 1 69 4 01 6 85 9 14 12 29 35 1 69 4 02 6 84 9 10 12 13 40 1 70 4 03 6 84 9 07 11 99 45 1 70 4 04 6 83 9 04 11 84 50 1 71 4 06 6 83 9 01 11 70 10 20 If the sub display does not automatically recognize a buffer it means the electrode could be defective or the buffer solution is wrong Calibration cannot proceed until they are rectified 10 21 When the READY sign appea...

Page 13: ...e points 10 33 At least 2 calibration points offset slope must be established in one session of calibration Otherwise erroneous reading will result 10 34 Only upon completing the entire session of calibration press to exit calibration mode and to begin measurement 11 MAKING PH MEASUREMENT pH measurement with ATC Automatic Temperature Compensation on sensor error 11 1 This meter comes with an elect...

Page 14: ...e not required during measurement 12 2 Rinse the sensor in distilled water then in the test liquid Stir to remove bubbles and wait for a stabilized reading 12 3 When the READY sign appears a stable reading has established 12 4 Depending on the liquid that is measured some volatile liquid may take as long as 20 minutes or even longer time to be stabilized Therefore even the READY sign appears the e...

Page 15: ...eading or slow response 1 Ground loop in solution make measurement in a cup or container 2 Electrode is clogged 3 Electrode could be expired 4 Electromagnetic interference move to another location to perform test Reading does not change 1 Broken pH glass sensor with micro crack 2 HOLD function is activated Press to release Wrong date time Backup battery has expired Contact your dealer to replace n...

Page 16: ... transinstruments com Singapore Pte Ltd email sales transinstruments com Website www transinstrumnets com ISO9001 Certified Firm Rev 4 July2014 PRINTED IN SINGAPORE Quality checked in Singapore CALIBRATION CERTIFICATE Product Professional Bench top pH meter BP3001 Measuring range 0 to 14 pH Calibration point s 7 00 4 01 10 01 Test point s 7 00 4 01 10 01 Accuracy achieved 0 02 Complete system mete...

Reviews: