Trans instruments BP3001 Operation Manual Download Page 13

12

 

Calibration with Custom Standard 

10.25 Make sure the unit is properly installed and electrode connected. 

10.26 Press  

 

and  display  will  show 

CAL

CAL

CAL

CAL

  flashing  on  the  left, 

indicating it is in calibration mode to make an Offset point calibration. 

10.27 Rinse the pH electrodes in distilled water then into the buffer solution 

for Offset calibration or the slope buffer on continuity. 

10.28 Main  display  will  show  the  current  measured  value  and  the  bottom 

sub-display will show 

2

2

2

2

.

00

00

00

00

 or the previous value on continuity.  

10.29 Press   

 

to  select  the  value  nearest  to  the  calibration  solution.  

The preset custom standards are 

2

2

2

2

.

00

00

00

00

4

4

4

4

.

50

50

50

50

7

7

7

7

.

00

00

00

00

9

9

9

9

.

50

50

50

50

 and 

12

12

12

12

.

.50

.50

.50

.50

10.30 After selecting the nearest value, press    

or 

 

to  adjust  to 

the exact value as the buffer solution at the current temperature being 
used.  

10.31 Press  

 

to confirm and display will flash then revert to raw data. 

10.32 Now  you  are  ready  to  make  Slope  point  calibration.    Repeat  step 

10.27 to 10.31 in sequence.  You can calibrate up to 4 slope points. 

10.33 

At  least  2  calibration  points  (offset  &  slope)  must  be  established  in  one 
session of calibration.  Otherwise, erroneous reading will result.

      

10.34 Only upon completing the entire session of calibration, press         to 

exit calibration mode and to begin measurement. 

 

11.  MAKING PH MEASUREMENT: 

pH measurement with ATC (Automatic Temperature Compensation on sensor error) 

11.1  This  meter comes  with  an  electrode  with  built  in  temperature sensor.  

Plug in the BNC and mini-phone connector as in section 6 of page4.  
The ATC sign will appear on the display. 

11.2  Remove the electrode bottle carefully as in 5.4 of page3.  Rinse probe 

in  distilled  water  before  each  test.    Shake  the  electrode  with  a  snap 
motion,  the  way  you  would  to  a  mercury  thermometer  and  dip  it  into 
the test solution, stir to remove any bubbles. 

11.3  Once  a  reading  is  stabilized  and  the 

READY

  sign  appear.    You  may 

record the measurement. 

Note on Measurements and Analysis:

 

Each  type  of  liquid  has  its  own  pH  value  varies  at  different  temperature.  
ATC  measurement  only  compensates  for  the  pH  sensor’s  error,  it  does  not 
compensate for the pH variation due to temperature fluctuation in the liquid.  
Therefore,  each  test  should  be  made  at  the  same  temperature  if  possible  or 
recorded with temperature reading for a comprehensive analysis. 

Summary of Contents for BP3001

Page 1: ...0 15 0 15 0 1 14 1 14 13 2 13 2 3 12 3 12 11 4 11 4 5 10 5 10 9 6 9 6 7 8 7 8 Operation Manual Professional Benchtop pH meter BP3001 ...

Page 2: ...ATURES Multiple display of pH temperature simultaneously Automatic buffer recognition with built in ISO and NIST standard Manual setting of customize buffer standard 5 points calibration Hold or freeze display function 99 memory with real time clock recording Maximum and minimum data review RS232 online data logging to PC 2 CONTENT Carefully unpack the box It should contain the following items a M...

Page 3: ...Operating humidity Up to 95 RH Meter size 217 x 168 x 58mm Package size Package weight 2 1kg 4 PRODUCT ASSEMBLY Turn anti clockwise to loosen arm to adjust angle turn clockwise to lock arm into position Place meter s left corner onto the STAND BASE to set into position Electrode stand can be set on the right corner of meter as desired Complete the stand assembly by pushing the arm into the top hol...

Page 4: ...ulb is half of the measuring element and must be kept wet to enhance speed of measurement If the sensor is dried out it will take a longer time to get hydrated The other measuring element is the reference element which contain concentrated chemical with a liquid bridge of porous fiber allowing controlled contact with the measuring liquid In order to keep both sensors in ideal condition the electro...

Page 5: ...is being saved MAX Maximum reading of all saved data displayed MIN Minimum reading of all saved data displayed REC Flashing while recall mode is initiated HOLD Display put on hold READY When the READY feature is enabled during measurement it will flash indicating measurement in progress When stopped indicates reading has stabilized 7 5 4 2 1 6 3 Socket for RS232 cable for PC data transfer AC DC Ad...

Page 6: ...CALL mode MODE KEY MOVE UP SELECT KEY 1 Press to switch display mode between pH and mV 2 During setting press to move digit up or next selection MEMORY SAVE KEY MOVE DOWN SELECT KEY 1 Press to save current reading into memory 2 During setting press to move digit down or previous selection CONFIRM KEY MAXIMUM MINIMUM DISPLAY KEY TEMPERATURE SET KEY 1 During CALIBRATION or SETUP setting press to con...

Page 7: ...he PC 4 Under Available ports select the com port meter is connected to 5 Click to select file location and type file name then click save 6 Set Baud rate to 9600 Data bits to 8 Parity to non Stop bits to 1 Flow control to non and Click 7 Press on the meter and display will show out out out out P1 P1 P1 P1 1 1 1 1 flashing PC will indicate transmission in progress 8 When transmission is completed ...

Page 8: ...lue not greater than 60mV Otherwise the electrode should be replaced 5 Press to return to ELE ELE ELE ELE P3 P3 P3 P3 0 0 0 0 6 Press to go to next parameter Buf P4 P4 P4 P4 0 0 0 0 BUFFER CALIBRATION STANDARD This setting allows you to select ISO NIST or customized buffer standard Factory preset is ISO standard ISO i i i i So So So So buffer standards are pH1 68 7 00 4 01 10 01 12 45 and NIST n n...

Page 9: ...ment for temperature Factory preset is in C 1 Press and display will show Unit Unit Unit Unit P7 P7 P7 P7 1 1 1 1 with C C C C flashing 2 Press to change between F F F For C C C C 3 Press to confirm setting display will return to Unit Unit Unit Unit P7 P7 P7 P7 0 0 0 0 4 Press to go to next parameter Rtc P8 P8 P8 P8 0 0 0 0 REAL TIME CLOCK This setting allow you to change the date in Year Month Da...

Page 10: ...de in the acidic range select 2 points between pH0 to 7 If measurements are made in the alkaline range then select between pH7 to 14 If full range is required calibrate at least 3 to 5 calibration points between pH0 to 14 10 2 Buffer standard must be set during SETUP in page 7 BuF BuF BuF BuF P4 P4 P4 P4 0 0 0 0 10 3 Standard solutions are in 3 groups of standards ISO Nist and custom standards ISO...

Page 11: ...01 10 01 12 45 30 1 69 6 98 4 02 9 97 12 29 35 1 69 6 98 4 02 9 93 12 13 40 1 70 6 97 4 03 9 89 11 99 45 1 70 6 97 4 04 9 86 11 84 50 1 71 6 97 4 06 9 83 11 70 10 10 If the sub display does not automatically recognize a buffer it means the electrode could be defective or the buffer solution is wrong Calibration cannot proceed until they are rectified 10 11 When the READY sign appear on the top lef...

Page 12: ...3 25 1 68 4 01 6 86 9 18 12 45 30 1 69 4 01 6 85 9 14 12 29 35 1 69 4 02 6 84 9 10 12 13 40 1 70 4 03 6 84 9 07 11 99 45 1 70 4 04 6 83 9 04 11 84 50 1 71 4 06 6 83 9 01 11 70 10 20 If the sub display does not automatically recognize a buffer it means the electrode could be defective or the buffer solution is wrong Calibration cannot proceed until they are rectified 10 21 When the READY sign appea...

Page 13: ...e points 10 33 At least 2 calibration points offset slope must be established in one session of calibration Otherwise erroneous reading will result 10 34 Only upon completing the entire session of calibration press to exit calibration mode and to begin measurement 11 MAKING PH MEASUREMENT pH measurement with ATC Automatic Temperature Compensation on sensor error 11 1 This meter comes with an elect...

Page 14: ...e not required during measurement 12 2 Rinse the sensor in distilled water then in the test liquid Stir to remove bubbles and wait for a stabilized reading 12 3 When the READY sign appears a stable reading has established 12 4 Depending on the liquid that is measured some volatile liquid may take as long as 20 minutes or even longer time to be stabilized Therefore even the READY sign appears the e...

Page 15: ...eading or slow response 1 Ground loop in solution make measurement in a cup or container 2 Electrode is clogged 3 Electrode could be expired 4 Electromagnetic interference move to another location to perform test Reading does not change 1 Broken pH glass sensor with micro crack 2 HOLD function is activated Press to release Wrong date time Backup battery has expired Contact your dealer to replace n...

Page 16: ... transinstruments com Singapore Pte Ltd email sales transinstruments com Website www transinstrumnets com ISO9001 Certified Firm Rev 4 July2014 PRINTED IN SINGAPORE Quality checked in Singapore CALIBRATION CERTIFICATE Product Professional Bench top pH meter BP3001 Measuring range 0 to 14 pH Calibration point s 7 00 4 01 10 01 Test point s 7 00 4 01 10 01 Accuracy achieved 0 02 Complete system mete...

Reviews: