MODEL
6535:
TraceableGO™
Datalogging Thermometer
6537:
TraceableGO™ Datalogging Hygrometer
Figure 1:
6535 (left), 6537 (right)
Dimension
L x H x D: 3.5 x 2 x 0.79” (89x51x20mm)
TEMPERATURE
Operating Range:
–20 to 70°C
Resolution:
0.1°C
Accuracy:
6535: ±0.4°C
6537: ±0.4°C between –10 and 70°C, otherwise
±0.5°C
RELATIVE HUMIDITY (6537 ONLY):
Range:
0% to 95%RH, non-condensing
Resolution:
0.1% RH
Accuracy:
±3% RH between 5 to 75%, otherwise
±5% RH
BATTERY
• 2 AAA Alkaline batteries (3.0V)
Battery Level Indication:
Battery level
LCD symbol
≥ 80% (2.78V)
≥ 60% (2.56V), < 80%
≥ 40% (2.34V), < 60%
≥ 20% (2.12V), < 40%
≥ 10% (2.01V), < 20%
< 10% Flashing
Note:
Battery level is updated every 5 minutes.
Note:
Once battery level drops to below 10%, device
may not work properly. Replace batteries immediately.
Note:
When replacing batteries, after removing the
old batteries, wait 10 seconds before inserting new
batteries. Otherwise, Bluetooth may not work properly.
MEASUREMENT READING UPDATE FREQUENCY
Temperature and humidity: 5 seconds;
Note:
If a reading is out of operating range, the
corresponding location on LCD will display ‘--.-‘, and
such out of range reading will not trigger alarm.
DATA LOGGING FREQUENCY:
5 min by default, user-adjustable between 1 minute
and 12 hours with 1 minute step.
DATA STORAGE CAPACITY
Alarm:
most recent 90 Alarm events
Data:
64K (65536) data points, 7.5 months on
5-minute logging interval
Note:
If maximum number of alarm events is reached,
device will overwrite the oldest alarm events.
_____________________________________________________________________________________
Page
3
of
10
Figure 2
Device not initialized
Figure 3
Device’s S/N not programmed
10.Device Operating Modes:
•
IDLE Mode
: First time battery inserted, and device has not been configured;
•
STANDBY Mode
: Device has been configured, but not started;
•
RUN Mode
: Device starts to measure and log data.
•
STOP Mode
: Device stops from RUN Mode. In STOP Mode, device does not update
measurements or log data, and the last measurements are displayed.
•
TEST Mode
: For test and calibration in production.
•
DTM Mode
: For Bluetooth radio test.
11.View Current Reading
1) 6535: Current reading, minimum/maximum reading since last clearing, run
time/alarm time toggling, Memory mode are shown on LCD.
2) 6537: Current temperature/humidity reading toggling every 5 seconds,
minimum/maximum reading since last clearing, run time/alarm time, Memory mode
are shown on LCD.
12.View Current MIN/MAX
6535: Current temperature Min/Max values are displayed on LCD
6537: Current temperature/humidity Min/Max values are toggled display.
Note: Every time device is configured, or resumes to run from STOP Mode, Min/Max
values are reset.
13.Viewing RUN TIME/ALARM TIME
Run Time/Alarm Time is toggled to display on LCD. If Run Time is displayed, LCD symbol
appears; if Alarm TIME is displayed, LCD symbol
appears.
Note: Alarm Time is accumulated for both Low Alarm and Hi Alarm for each channel
(temperature, humidity).
14.View Memory Usage, Current Date/Time
, Device’s Serial Number
1) Short press and release
;
DEVICE OPERATING MODES
•
IDLE Mode: First time battery inserted, and device
has not been confi gured;
•
STANDBY Mode: Device has been confi gured, but
not started;
• RUN Mode: Device starts to measure and log data.
• STOP Mode: Device stops from RUN Mode.
In STOP Mode, device does not update
measurements or log data, and the last
measurements are displayed.
VIEW CURRENT READING
1. Temperature ONLY unit: Current reading,
minimum/ maximum reading, run time/alarm time
toggling, Memory mode are shown on LCD.
2. Humidity and Temperature unit: Current
temperature/humidity reading toggling every 5
seconds, minimum/maximum reading since last
clearing, run time/alarm time, Memory mode are
shown on LCD.
VIEW CURRENT MIN/MAX
Temperature ONLY unit: Current temperature Min/Max
values are displayed on LCD.
Humidity and Temperature unit: Current temperature/
humidity Min/Max values are toggled display.
Note:
Every time device is confi gured, or resumes to
run from STOP Mode, Min/Max values are reset.
VIEWING RUN TIME/ALARM TIME
Run Time/Alarm Time is toggled to display on LCD.
If Run Time is displayed, LCD symbol
RUN TIME
appears; if Alarm Time is displayed, LCD symbol
ALARM TIME
appears.
Note:
Alarm Time is accumulated for both Low
Alarm and Hi Alarm for each channel (temperature,
humidity).
VIEW MEMORY USAGE, CURRENT DATE/TIME,
DEVICE’S SERIAL NUMBER
1. Press and release
START/STOP
button
2. Memory usage in percentage is shown on LCD.
The percentage indicates how much internal data
memory storage has been used;
3. Number of days, hours, minutes left till memory is
full are also displayed on the second line;
_____________________________________________________________________________________
Page
4
of
10
2) Memory usage in percentage is shown on LCD. The percentage indicates how much
internal data memory storage has been used;
3) Number of days, hours, minutes left till memory is full are also displayed on the
second line;
Figure
3
4) Within 10 seconds, short press
again, current date/time is displayed on device
LCD. The following figure shows 9/14/2017, 17:30
Figure
4
5) Within 10 seconds, short press
again, device S/N will be displayed on LCD.
Figure
5
6) To go back to normal working status, press
again, or wait for 10 seconds and
device will automatically go back to normal working status.
Note:
If Memory Mode is set to WRAP WHEN FULL: LCD symbol
starts to
flash on the display when the memory is full. Once memory is full, the oldest data
points will be overwritten with new data points.
4. Within 10 seconds, press
START/STOP
again,
current date/time is displayed on device LCD. The
following fi gure shows 9/14/2017, 17:30
_____________________________________________________________________________________
Page
4
of
10
2) Memory usage in percentage is shown on LCD. The percentage indicates how much
internal data memory storage has been used;
3) Number of days, hours, minutes left till memory is full are also displayed on the
second line;
Figure
3
4) Within 10 seconds, short press
again, current date/time is displayed on device
LCD. The following figure shows 9/14/2017, 17:30
Figure
4
5) Within 10 seconds, short press
again, device S/N will be displayed on LCD.
Figure
5
6) To go back to normal working status, press
again, or wait for 10 seconds and
device will automatically go back to normal working status.
Note:
If Memory Mode is set to WRAP WHEN FULL: LCD symbol
starts to
flash on the display when the memory is full. Once memory is full, the oldest data
points will be overwritten with new data points.
5. Within 10 seconds, press
START/STOP
again,
device S/N will be displayed on LCD.
_____________________________________________________________________________________
Page
4
of
10
2) Memory usage in percentage is shown on LCD. The percentage indicates how much
internal data memory storage has been used;
3) Number of days, hours, minutes left till memory is full are also displayed on the
second line;
Figure
3
4) Within 10 seconds, short press
again, current date/time is displayed on device
LCD. The following figure shows 9/14/2017, 17:30
Figure
4
5) Within 10 seconds, short press
again, device S/N will be displayed on LCD.
Figure
5
6) To go back to normal working status, press
again, or wait for 10 seconds and
device will automatically go back to normal working status.
Note:
If Memory Mode is set to WRAP WHEN FULL: LCD symbol
starts to
flash on the display when the memory is full. Once memory is full, the oldest data
points will be overwritten with new data points.
6. To return to normal working status, press
START/
STOP
again, or wait for 10 seconds and device will
automatically go back to normal working status.
Note:
If Memory Mode is set to WRAP WHEN FULL:
LCD symbol
_____________________________________________________________________________________
Page
4
of
10
2) Memory usage in percentage is shown on LCD. The percentage indicates how much
internal data memory storage has been used;
3) Number of days, hours, minutes left till memory is full are also displayed on the
second line;
Figure
3
4) Within 10 seconds, short press
again, current date/time is displayed on device
LCD. The following figure shows 9/14/2017, 17:30
Figure
4
5) Within 10 seconds, short press
again, device S/N will be displayed on LCD.
Figure
5
6) To go back to normal working status, press
again, or wait for 10 seconds and
device will automatically go back to normal working status.
Note:
If Memory Mode is set to WRAP WHEN FULL: LCD symbol
starts to
flash on the display when the memory is full. Once memory is full, the oldest data
points will be overwritten with new data points.
starts to fl ash on the display when
the memory is full. Once memory is full, the oldest
data points will be overwritten with new data points.
If Memory Mode is set to STOP WHEN FULL: LCD
symbol
MEM
starts to fl ash on the display when the
memory is 95% full. Once memory is full, the device
will stop logging new data points.
ENABLE/DISABLE BLUETOOTH
1. Quick press twice
START/STOP
to enable
Bluetooth, and Bluetooth LCD symbol appears;
2. Device starts to advertise to be found, and
to be connected. Device Name shown on
TraceableGO™ App discovered list looks like
CC6535-xxxx, or CC6537-xxxx, where xxxx is the
last 4 digits of device serial number.
3. If no connection has been made for
ONE MINUTE
,
Bluetooth will be disabled to save battery life, and
Bluetooth LCD symbol disappears, or quick
press twice again to disable Bluetooth. To enable
again, go to Step (1).
HOW TO SETUP DEVICE
Confi gure device before use
1. Enable Bluetooth on device.
2. Use TraceableGO™ App to connect to device.
3.
Follow TraceableGO™ App instruction to confi gure
device.
4.
The following parameters are confi gured through
TraceableGO™ App: START Mode, STOP Mode,
ALARM Enable/Disable, Celsius/Fahrenheit,
Memory Mode, Data Logging Interval, Alarm
Setting.
5.
Once device has been confi gured, device enters
STANDBY Mode.
Note:
While device is connected to our Traceable-
GO™ App, if no action has been taken for 5 minutes,
device will terminate the connection from our
TraceableGO™ App.
Note:
If Memory Mode is set to WRAP WHEN FULL,
_____________________________________________________________________________________
Page
4
of
10
2) Memory usage in percentage is shown on LCD. The percentage indicates how much
internal data memory storage has been used;
3) Number of days, hours, minutes left till memory is full are also displayed on the
second line;
Figure
3
4) Within 10 seconds, short press
again, current date/time is displayed on device
LCD. The following figure shows 9/14/2017, 17:30
Figure
4
5) Within 10 seconds, short press
again, device S/N will be displayed on LCD.
Figure
5
6) To go back to normal working status, press
again, or wait for 10 seconds and
device will automatically go back to normal working status.
Note:
If Memory Mode is set to WRAP WHEN FULL: LCD symbol
starts to
flash on the display when the memory is full. Once memory is full, the oldest data
points will be overwritten with new data points.
appears on LCD; If Memory Mode is set to
STOP WHEN FULL,
MEM
appears on LCD.
ALARM
1. Once an alarm is triggered by any of temperature,
humidity out of set alarm range, LCD symbol
LOW
ALM
and/or
HI ALM
starts to fl ash. An Alarm Event
will be logged.
2. Pressing
START/STOP
once will clear the alarm,
LCD symbol stops fl ashing. An Alarm Acknowledge
Event will be logged.
3. If temperature, or humidity falls back to normal
range, an Alarm Event will be logged. If any of
temperature, humidity measurement goes back to
out of alarm range, alarm will be triggered again.
HOW TO START, STOP, AND RESUME DATA
LOGGING
1. Devices are factory initialized with default setting:
• START Mode: Push Start
• STOP Mode: Push Stop
•
Memory Mode: Wrap when Memory is Full
• Temperature Unit: Celsius
• Data Logging Interval: 5 minutes
• Alarm Enable/Disable: Enabled
• Alarm Setting High Alarm:
▪ Temperature: 8°C
▪ Humidity: 75% RH (6537 only)
• Alarm Setting Alarm Low:
▪ Temperature: 2°C
▪ Humidity: 25% RH (6537 only)
• Date/Time is also set to current to central time.
Note:
Devices are shipped with battery, powered
on, and ready to use if using default setting. Without
powered on, device will lose track of date/time.
Note:
If a user intends to use device for different
setting other than default, device must be re-
initialized.
_____________________________________________________________________________________
Page
1
of
10
6
535
/
6
537
B
LUETOOTH
D
ATA
L
OGGER
I
NSTRUCTION
11/01/2017
1.
Mode Number:
6535: Bluetooth Data Logging Thermometer
6537: Bluetooth Data Logging Hygrometer
2.
Device Display:
LCD View Angle: 6
O’clock
Figure 1:
6535 (left), 6537 (right)
Dimension L x H x D:
3.5” x 2” x 0.
79
” (
89mm x 51mm x 20mm)
3. Temperature:
Operating Range: -20°C to 70°C
Resolution: 0.1°C
Accuracy:
6535: ±0.4°C
6537: ±0.4°C between -10°C and 70°C, otherwise ±0.5°C
4. Relative humidity (6537 Only):
Range: 0% to 95%RH, non-condensing
Resolution: 0.1% RH
Accuracy: ±3% RH between 5 to 75%, otherwise ±5% RH
Recommended Humidity Operating Range: 20% to 80% (non-condensing) over -10°C to 60°C.
5. Battery
•
2 AAA Alkaline batteries (3.0V)
•
Battery Level Indication:
Battery Level
LCD Symbol
≥
80% (2.78V)