SPECIFICATIONS  

Backlight Feature

Press the         button and the backlight illuminates for approxi-
mately 20 seconds then turns off automatically.

Measurement Modes

The meter can be set to measure pressure in the following 
modes: bar, mmHg, ozin

2

, kgcm

2

, psi, inH

2

0, kPa, ftH

2

0, inHg, 

cmH

2

0 and mbar.  Each press of the UNIT button will cycle the 

meter through each of the measurement modes.  An arrow at 
the bottom of the LCD will indicate the mode selected.  The 
meter will default to the mode last selected before meter was 
turned off.

Clearing Display to Zero

To clear the display to zero disconnect from all pressure and 
press the HOLD button for 3 seconds. “o o o o” will appear on 
the LCD and the meter value will be cleared to zero.  

Hold Mode

Pressing the HOLD button once will “freeze” the measurement 
reading. “HOLD” will appear at the upper left of the LCD indi-
cating the meter is in the hold mode.  Press the HOLD button 
again to return the meter to pressure measurement mode.  

Differential Mode

Differential pressure is the measurement of the difference 
between two pressures.  

TRACEABLE

®

 MANOMETER/PRESSURE/VACUUM INSTRUCTIONS

(Continued on next page )

Display:  

1

/

2

", 4- digit LCD.

Accuracy:  

±0.3% of full scale between 0 to 50

o

C

Size/Weight: 

2

3

/

4

 x 7

1

/

8

 x 1

1

/

8

  / 5 oz

Response time: 

0.5 seconds

Operating

 

Temp

 

Range:

  32-122

o

F (0-50

o

C)

Features: Memory 

feature,

 

Backlight display, 

 

RS-232 capability.

CONTROL BUTTONS

 

Powers unit off and on

REC   Record mode
HOLD   "Freeze" the pressure reading on the display
DIF -   Differential mode
UNIT Cycle 

meter  through pressure units of measurement 

 

Backlight function

OPERATION
On / Off / Auto Power Off

Press the          button to turn the meter on. The meter will shut 
off automatically approximately 20 minutes after every power on.  
The “auto power off” function can be disabled for recording or 
operating over longer periods of time by pressing POWER and 
HOLD buttons simultaneously before power on. An “n” will appear 
in the middle of the screen at which time the HOLD button can be 
released, and then the power button. The “auto power off” mode 
has now been disabled. 

RANGE AND RESOLUTION CHART

  Cat. No. 

3460 3461 3462 3463 

  Maximum Allowed PSI 

20 30 60 150

  PSI Range 

–5.000 to 5.000 

–15.00 to 15.00 

–30.00 to 30.00 

–100.0 to 100.0

  PSI Resolution 

0.003 0.01  0.02 0.1

  mbar Range 

–345.0 to 345.0 

–1034 to 1034 

–2068 to 2068 

–6895 to 6895

  mbar Resolution 

0.2 1  2 4

  bar Range 

–0.3450 to 0.3450 

–1.034 to 1.034 

 –2.068 to 2.068 

–6.895 to 6.895

  bar Resolution 

0.001 0.001 0.002 0.004

  cm/H

2

O Range 

–351.0 to 351.0 

–1055 to 1055 

–2109 to 2109 

–7031 to 7031

  cm/H

2

O Resolution 

0.2 1  2 4

  inch/H

2

O Range 

–138.0 to 138.0 

–415.0 to 415.0 

–830.0 to 830.0 

–2768 to 2768

  inch/H

2

O Resolution 

0.1 0.3 0.5  2

  feet/H

2

O Range 

–12.00 to 12.00 

–35.00 to 35.00 

–69.00 to 69.00 

–231.0 to 231.0

  feet/H

2

O Resolution 

0.01 0.02 0.04  0.2

  mm/Hg Range 

–259.0 to 259.0 

–776.0 to 776.0 

–1551 to 1551 

–5171 to 5171

 

  mm/Hg Resolution 

0.2 0.5  1 3 

  inch/Hg Range 

–10.00 to 10.00 

–31.00 to 31.00 

–61.00 to 61.00 

–204.00 to 204.00

  inch/Hg Resolution 

0.01 0.03  0.1 1.0 

  kg/cm

2

 Range 

–0.350 to 0.350 

–1.060 to 1.060 

–2.110 to 2.110 

–7.030 to 7.030

  kg/cm

2

 Resolution 

0.001 0.001 0.002 0.004 

  oz/inch

2

 Range 

–80.00 to 80.00 

–240.0 to 240.0 

–480.0 to 480.0 

–1600 to 1600

  oz/inch

2

 Resolution 

0.05 0.2  0.3 1

  kPa Range 

–34.50 to 34.50 

–103.0 to 103.0 

–207.0 to 207.0 

–690.0 to 690.0

  kPa Resolution 

0.02 0.1  0.2 0.4

Reviews: