Metalyser
®
Deluxe HM2000
Instruction Manual v2.0
Altering deposition times
The standard default deposition time of 60 seconds is suitable for all analysis
down to 5ppb for most elements, 10ppb for arsenic. To increase accuracy for the
lower levels, an increased deposition time is recommended. If the user expects
the analysis to be higher, then a shorter deposition time should be used. This can
be adjusted when prompted to choose a deposition time at the beginning of the
test methods.
30, 60 or 120 seconds can be selected by moving the joystick to the right and to
the left on
this screen and then pushing the centre of the joystick to accept it.
Deposition time (seconds)
Concentration range (ppb)
30
50 to 500
60
10 to 50
120
<10
Suggested ranged for choosing deposition times
The instrument has been designed to have a maximum reading of 500ppb. Should the
user wish to measure higher than this they will need to dilute the sample. For example, if
the user wishes to measure 600ppb dilute the sample by half and then multiply the result
by 2.
Always dilute with water that will not contribute to the heavy metals being
tested. Deionised water is ideal.
Blank subtraction
Blank subtraction is recommended to be used for all of the analyses and is the default
setting. It is particularly important for the analysis of Hg and As.
Summary of Contents for METALYSER DELUXE HM2000
Page 1: ...METALYSER DELUXE HM2000 Portable Heavy Metals Analyser...
Page 2: ...2 www trace2o com Metalyser Deluxe HM2000 Instruction Manual v2 0 CONTENTS...
Page 37: ...www trace2o com 37 Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...
Page 38: ...38 www trace2o com Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...
Page 39: ...www trace2o com 39 Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...