8.15
INSTRUMENT SETUP – OPTICS ADJ.
The unique design of the TQC Automatic Cupping Test allows
the user to use optional available loupe or USB microscope.
The latter has a limited depth of focus and needs to be
adjusted to zero level. In its base position the indenter is about
0.5mm below the test panel. An USB microscope can only be
adjusted when the indenter is against the test panel. In order
to bring the indenter to its zero position the optical
adjustment menu can be used.
When the TQC Automatic Cupping Test informs you to adjust
the optical tool you are able to adjust the camera. Always
adjust the camera on the panel you are planning to do your
tests on. The thickness of a panel and coating will influence
your focus. Select back for the indenter to move back to their
base position.
In the base position both the indenter and the camera focus
are 0.5mm below the panel. The camera support is not fixed
to one position. It moves along with the indenter to keep
your camera focussed on the top of the indentation,
allowing you to always have a perfect picture.
8.16
RUN
When performing a run the displayed screens vary depending
on the in “Run setup” selected type of run. Both run menu
types will be explained in the next two paragraphs.
8.17
PRESET RUN
In the first screen the TQC Automatic Cupping Test asks the
user to place the test panel and close the clamp. When done
select confirm to continue.