background image

2 |

2

 

WHAT’S IN THE BOX? 

Ultrasonic wall thickness gauge 
Coupling gel 60ml 
Probe 10mm - 2.5MHz 
Probe 10mm - 5.0MHz 
Calibration plate 

3

 

OPERATION INSTRUCTIONS 

š

 

Operation area condition: This gauge can measure all samples which are equal to or bigger than the probes 
surface area.  

š

 

Curved surface condition: When materials have a curved surfaces like a boiler wall or tubing, the curvature 
radius must be equal to or more than 10mm and the wall thickness must be equal to or more than 3mm.  

š

 

Roughness condition: This gauge is applicable for a wide range of materials. In most cases the supplied 
probe makes very accurate readings. However if the roughness is too big due to rust etc. an error may 
occur. In this case please try to minimise the roughness or select the 2.5MHz probe. If the probe is worn-
out, please contact your local distributor for a replacement probe. 

š

 

Working temperature condition: Material and sound velocity will change along with temperature. In 
normal measurement conditions environment temperature impact can be ignored. The probe is made of 
propylene material. Considering the protection of the probe and its precision we recommend a surface 
temperature of the work piece should not be over 60

o

C otherwise the probe could be damaged. 

š

 

Operation temperature: 0 - 40

o

C  

š

 

Relative humidity: <90% RH 

š

 

Hardware / Work piece/ Material temperature: < 60

o

š

 

Do not apply on heavily vibrating / erosive material 

š

 

Avoid impact and moisture 

4

 

PREPARATIONS 

1. Turning on your instrument

Connect the probe with the main unit, press ON/OFF to turn on the instrument. The display shows a full screen 
for 0.5 sec with backlight, then a vertical bar is displayed twice circularly, meanwhile the main display area 
shows 0000 to 9999 progressively. After that  the LCD displays the last applied sound velocity with registered 
memory.  

 

Full screen 

Last applied sound velocity & unit no. 

Summary of Contents for LTD7015

Page 1: ...will give accurate readings to an accuracy of 1 This unit is not suitable for cast iron due to its big crystalloid composition 1 1 Specifications Measuring method Ultrasonic pulse echo Measuring freq...

Page 2: ...ng temperature condition Material and sound velocity will change along with temperature In normal measurement conditions environment temperature impact can be ignored The transducer is made of propyle...

Page 3: ...eted After calibration the sound velocity will remain the same as previous selected A calibration should be made the every time the probe or batteries are replaced This is very important to make sure...

Page 4: ...When the icon flashes please replace the batteries before taking any new measurements Turn the instrument off by pressing the ON OFF button Open the battery lid and replace them with 3 pcs of AAA alk...

Page 5: ...or otherwise is correct we have no control over either the quality or condition of the product or the many factors affecting the use and application of the product Therefore unless we specifically agr...

Page 6: ...e for cast iron due to its big crystalloid composition 1 1 Specifications Measuring method Ultrasonic pulse echo Measuring frequency 5MHz 2 5MHz Measuring range Accuracy Sound velocity Display Resolut...

Page 7: ...utor for a replacement probe Working temperature condition Material and sound velocity will change along with temperature In normal measurement conditions environment temperature impact can be ignored...

Page 8: ...small amount of coupling gel on the sample block and then place the probe on the sample block When the display shows 4 0mm the calibration is completed After calibration the sound velocity will remain...

Page 9: ...ss MODE again to store After selecting your memory unit you can start measuring again While measuring the memory unit will flash and the new measurement will delete the stored value When your measurem...

Page 10: ...the axis of the pipe to be measured For small pipes two measurements should be performed one with the axis of the pipe perpendicular the other one parallel The smallest of the 2 values displayed shou...

Page 11: ...calibrating the gauge with a reference block However this table is just for reference sometimes the sound velocity will be different caused by different physical chemical factors The sound velocity of...

Page 12: ...ve recess surface incorrect calibration sample blocks or the inner flaw of materials 8 Coupling gel Coupling gel causes the high frequency ultrasonic waves transmitting between the probe and material...

Page 13: ...t in use We recommend annual calibration 9 DISCLAIMER The right of technical modifications is reserved The information given in this manual is not intended to be exhaustive and any person using the pr...

Reviews: