15
indicates that a batch was previously used for manual
logging.
5. If required, select
Limits
and press OK to set the limits for
the selected batch and indicate whether the limits should
be used (see Section 3.2 for details on setting limits).
Select
Back
and press OK to return to the
Batch
menu
and start logging.
Figure 4-10 Limits in batch
6. Place the rubber tip of the surface temperature sensor
against the surface at an angle of 90° (see figure 4-11),
and keep it in this position until the measured value
remains stable.
DO NOT USE EXCESSIVE FORCE!
A light contact with the surface is sufficient. Using force
does not increase the measuring speed or improve the
accuracy of the measurement, but it does wear and/or
damage the sensor tip. Do not drag the probe over the
surface.
Since holding the surface temperature sensor against a
surface for longer periods of time is impractical, we
recommend using the optional DewCheck4 Dock, which is
equipped to permanently position the instrument against a
surface. See Section 5 for more information about the
DewCheck4 Dock.