11.7 GLP Data Format
GLP information is returned as 5 or 6 lines terminated by a carriage
return. When using the “?G” command (section 11.5), the computer must
respond with a character after receiving each line. For example…
WP-82Y V2.0 S1234 @ 31/12/10 12:00
Oxygen Zero= 0.0% @ 31/12/10 11:00
Oxygen Span= 100.0% @ 31/12/10 11:10
Oxygen Altitude= 5000m @ 31/12/10 11:20
Oxygen Pressure= 1013HPa @ 31/12/10 11:20
Temperature Offset= 1.0oC @ 31/12/10 11:30
ENDS
NOTE:
Either Altitude
OR
Pressure is sent, depending upon which was
selected the last time the instrument was calibrated. If the
Altitude or Pressure Correction was switched off at the time of
the last calibration, neither is displayed.