Page 22
8.2 Failed Calibration
If calibration has failed, the GLP function will reset the date and time to
zero. The
WP-82Y
still shows the results of the last successful
calibration. For example…
Oxy Zero 1.0%
@ 00/00/00 00:00
Oxy Span 100.0%
@ 00/00/00 00:00
Altitude 5000m
@ 00/00/00 00:00
Pressure 1013HPa
@ 00/00/00 00:00
Temp Offset 1.0
O
c
@ 00/00/00 00:00
Note that these calibration values are still used if further measurements
are taken without recalibrating.
8.3 Printing GLP Information to the RS232 Port
The GLP information stored in the instrument’s memory can be sent to a
printer or PC via the RS232 port.
1. Switch the meter on.
2. Ensure that the
WP-82Y
RS232 cable is connected to the instrument
and to the printer or PC.
3. Send the GLP information to the RS232 port.
(
F4:Setup
F3:GLP
F3:Print
)
4. The GLP information is sent to the RS232 port in formatted ASCII
text. For example…
WP82Y V2.0 S1234 @ 31/12/10 12:00
Oxygen Zero= 0.0% @ 31/12/10 11:00
Oxygen Span= 100.0% @ 31/12/10 11:10
Oxygen Altitude= 5000m @ 31/12/10 11:20
Oxygen Pressure= 1013HPa @ 31/12/10 11:20
Temperature Offset= 1.0oC @ 31/12/10 11:30
ENDS
NOTE:
Either Altitude
OR
Pressure is sent, depending upon which was
selected the last time the instrument was calibrated. If the
Altitude or Pressure Correction was switched off at the time of
the last calibration, then neither is displayed.