Page 8
2. Operating Modes
The function switch is used to select the required mode. The four
operating modes available are…
1.
Set Upper Limit
: Switch to
Set Upper Limit
when setting the
Salinity/TDS value at which the High LED and
relay output are activated. The relay output is
not operational in this mode.
Factory-set to activate when the reading is
above
the Upper Limit. Can be user set to be a
activate when the reading is
below
the Upper
Limit. See section 3.5.
2.
Set Lower Limit
: Switch to
Set Lower Limit
when setting the
Salinity/TDS value at which the Low LED and
relay output are activated. The relay output is
not operational in this mode.
Factory-set to activate when the reading is
below
the Lower Limit. Can be user set to be a
activate when the reading is
above
the Lower
Limit. See section 3.5.
3.
Calibrate
: Switch to Calibrate mode to place the
pro
CHEM
-S
into standby (eg. when
calibrating). This mode disables the two output
relays to avoid chemicals being added when
placing the sensor into calibration buffers.
4.
Automatic Control
: Switch to
Automatic Control
to enable the
Alarm LED’s and relay outputs. These will be
activated when the Salinity/TDS is outside the
Upper and Lower limit values. See section 3.5.