
Page 22
10.3 pH and mV Troubleshooting
Symptom
Possible Causes
Remedy
Unit fails to
calibrate, even with
new probe.
Calibration settings outside of
allowable limits due to
previous failed calibration.
Initialise the unit. See section
9.
1 Point calibration
fails (Asymmetry is
greater than +/-1.00
pH).
1.
Reference junction
blocked.
2.
Reference electrolyte
contaminated.
Clean reference junction, as per
instructions supplied with the
electrode.
Flush with distilled water and
replace electrolyte.
2 Point calibration
fails (Slope is less
than 85.0%).
1.
Incorrect primary buffer.
2.
Glass bulb not clean.
3.
Electrode is aged.
4.
Connector is damp.
5.
Buffers are inaccurate.
Ensure that you are using the
buffers which the
MC-81
has
been set to automatically
recognise (See section 8).
Clean glass bulb as per
instructions supplied with the
electrode.
Attempt rejuvenation, as per
instructions supplied with the
electrode. If not successful,
replace electrode.
Dry in a warm place.
Replace buffers.
Unstable readings.
1.
Reference junction
blocked.
2.
Glass bulb not clean.
3.
Bubble in glass bulb.
4.
Faulty connection to meter.
5.
Reference junction not
immersed.
6.
KCl crystals around
reference junction, inside
the electrolyte chamber.
Clean reference junction, as per
instructions supplied with the
electrode.
Clean glass bulb as per
instructions supplied with the
electrode.
Flick the electrode to remove
bubble.
Check connectors. Replace if
necessary.
Ensure that the bulb AND the
reference junction are fully
immersed.
Rinse electrolyte chamber with
warm distilled water until
dissolved. Replace electrolyte.
Continued next page...