Page 20
10.2 Conductivity and Salinity Troubleshooting
Symptom
Possible Causes
Remedy
Unit attempts Span
calibration instead
of Zero calibration.
Electrode has Zero error.
Thoroughly rinse electrode in
distilled water and allow to
completely dry in air before
attempting zero calibration.
If instrument does not calibrate
at Zero with electrode
disconnected, then the
instrument is faulty.
Standard calibration
fails, and span is
less than 75%.
1.
Electrode is not immersed
deeply enough.
2.
Electrode may have a
build-up of dirt or oily
material on electrode wires.
3.
Platinum-black coating has
worn off.
4.
Standard solution is
inaccurate.
5.
Electrode is faulty.
Immerse electrode at least to
the vent hole in the white
plastic cover.
Clean electrode, as per the
instructions detailed in section
12.4.
Electrode requires
replatinisation.
Return to the factory, or see
details in section 12.5.
Replace standard solution.
Return electrode to factory for
repair or replacement.
Standard calibration
fails, and span is
greater than 133%.
1.
White protective cover is
not fitted.
2.
Standard solution is
inaccurate.
3.
Electrode may have a
build-up of conductive
material, such as salt.
4.
Electrode is faulty.
The white protective cover
MUST be fitted for correct
readings.
Replace standard solution.
Clean electrode, as per the
instructions detailed in section
12.4.
Return electrode to factory for
repair or replacement.
Continued next page...