  In percent weigh mode , if you want to reset the percent precision and radix percent value , please press button

MODE

to reset .   

Exit percent weigh mode 

long press

MODE

],

choose the normal weigh mode 

NORMALg

 

 

                                                                                                                                               

Density Weigh Function 

 

Aim 

By  using  this  procedure  user  can  measure  the  density  value  for  solid  or  liquid  so  that  judge  the  sample 
desirable or not. (need to buy the necessary weighing kits from our company) 

Weighing method for solid   

Step 1

Use the density measure device to measure the solid density in the air; 

Step 2

Put the solid  into a certain liquid to measure the density.(The density of the certain  liquid must be 

known in advance)   

button

command

                                    steps                                                                    screen display                   

Long press

MODE

  Enter the choice of mode                         

                                1. Choose density mode                                     

DENSITY     

                Solid density 

             

                                   

SOLID

    

Short press

CAL

2. Confirm the solid density measure mode 

                                          Enter the liquid temperature coefficient settings     

01.00000

      d 

  The user can set the liquid temperature coefficient value on his own , please refer to Form1. 

 

Press

PRINT

to change the position

press

TARE

to scroll down the number , press

CAL

  to confirm , enter the solid density measure 

situation .   

                          3

The balance reminding to weigh in the air 

      

HI

 

                       0.0000 

g

 

                          4

Put the sample and weigh it in the air                     

HI

 

                                     

e.g.

The quality of the sample in the air is 6.8135g

         

6.8135 

 

button

command

                                    steps                                                                    screen display                   

 

short press

CAL

                                5. To record the current weigh value in the air           

LO

 

                                                                      and indicating solid weighed in liquid                     

6.8135 

g

 

                        6. Take off the solid                                                   

LO

 

                                It indicating solid weighed in liquid                         

0.0000 

g

 

                        7. Put the solid in liquid                                             

LO

 

                                                             

e.g. Quality of the solid is 1.3518g in liquid

                             

1.3518 

g

 

Short press

CAL

                              8. It records the weigh value in liquid                               

OK               

                                  In the meantime calculate the solid density 

                     

1.2398 

g

d

 

                                                            Measure result : 

1.2398 

g/cm

3

 

           

 

Summary of Contents for FPA Series

Page 1: ... series Electronic Balance Operation Manual TPS SCALES VIET NAM JSC THINH PHAT SCALES JOINT STOCK COMPANY 57 D1 Street Ward 25 Binh Thanh District HCMC Vietnam TEL 0084 28 62 888 666 FAX 0084 28 3512 7966 ...

Page 2: ...ghing Operation 3 Basic Weighing Function 3 Calibrate Adjust 4 Application Function 4 Counting Function 5 Percentage Weigh Function 5 Density Weigh Function 6 Operation Menu 8 Communication Parameters 9 Unit Conversion 10 Technical Specification 10 Repairing and Maintenance 12 ...

Page 3: ...gh precision sensor technology magnetic balance it is the ideal device for fast accurate weighing widely used in medical research institute schools enterprise road construction and defense department etc LCD backlight display easy and clear Operator Highly smart tare and back to zero within the whole weighing scope overload display Multi mode choice counting percentage weigh density weigh With RS ...

Page 4: ...e shaking or vibrating plate to make the balance stable Recommend to use aseismic marble Avoid to put balance in the place with big temperature change or air flow such as the place with direct sunlight or cold air outlet Please use the separate power socket to avoid the interruption by other device Please do not put anything on the scale when power on the balance Make balance adjust itself to the ...

Page 5: ...aration Connect the power supply when the panel display press button ON OFF to switch on the balance Pre heat Time In order to ensure the precision of weighing please preheat the balance completely reference to technical parameter form page 19 20 then start the following operation Example Basic weigh operation balance already pre heated we take 210g 0 0001g balance as an example button command ste...

Page 6: ...dicated scale Display the calibrated scale value after 5 sec 200 0000 g 4 Take off the scale complete calibration 0 0000 g Application Function Operation Choose the Mode button command steps panel display Long press MODE Enter the choice for mode Normal weigh mode NORMAL g Count mode COUNT pcs Percent weigh mode PERCENT Density measure mode DENSITY d Density measure mode is an option if you need p...

Page 7: ...x number 20 pcs Remove the samples 0 pcs Weigh the unknown numbers in this e g 200pcs 200 pcs In Count Mode if you want to reset the radix number value please press MODE to reenter the Count Mode Exit the Count Mode Long press MODE choose the normal weigh mode NORMALg Percent Weigh Function The end user can make a certain weight as definition 100 then make other weight displayed as the percent of ...

Page 8: ... SOLID d Short press CAL 2 Confirm the solid density measure mode Enter the liquid temperature coefficient settings 01 00000 d The user can set the liquid temperature coefficient value on his own please refer to Form1 Press PRINT to change the position press TARE to scroll down the number press CAL to confirm enter the solid density measure situation 3 The balance reminding to weigh in the air HI ...

Page 9: ... 13 0 99941 27 0 99654 The weigh method for liquid density Use density measure device the volume of planed standard sample must be known The user need to enter the volume of standard sample manually The latest data entered will be stored for further use Step 1 To measure the liquid density please weigh the standard sample in the air Step 2 Then weigh the standard sample in the liquid want to be me...

Page 10: ...s TARE ZERO at the same time in 3 seconds it displays Cod0000 Press PRINT to change the position press TARE to scroll the numbers or choose the parameters press ZERO to confirm and enter the next step Code Mark display Description Setting range Cod0001 Zero x x Zero range 0 0d to 6 0d Stdy x x Start sensitive range 0 0d to 6 0d Sens X Inner sensitivity grade 1 2 3 4 5 6 Filt X Filter radix 1 2 3 4...

Page 11: ...by the operator Blank 3 2Common Data Format ST 0000 0000 g CR LF LF ST Steady 0 Value 0000 0000 g CR LF US Unsteady 0 Value Number Display Area Unit Display Area Carriage Return Line Feed 4 The list of command which the communicating device can send in software exchange mode ASCII code No Command Function 1 Sample 2 P Print 3 T Tare 4 Z Back to Zero 5 U Unit Unit Conversion Mark Unit Conversion ra...

Page 12: ...evel Weigh scope 0 100g 0 110g 0 120g 0 200g 0 210g 0 310g value of scale division 0 0001g Standard weights value 100g 100g 100g 200g 200g 300g Repeatable tolerance standard tolerance 0 0003g Linear tolerance 0 0003g Stable time 4s Pre heat time 60 120 min Operation temperature 20 2 5 Pan diameter 90mm Overall dimensions 325 x 205 x 305mm L W H Weigh chamber dimensions 180mmX175mmX200mm L W H Net ...

Page 13: ...3100g 4100g 5100g 6100g Precision level Weigh scope 0 1100g 0 2100g 0 3100g 0 4100g 0 5100g 0 6100g value of scale division 0 01g Standard weights value 1000g 2000g 3000g 4000g 5000g 5000g Repeatable tolerance standard tolerance 0 02g Linear tolerance 0 02g Stable time 3s Pre heat time 60 90 min Operation temperature 20 7 5 Pan diameter ф160mm Overall dimensions 325mm 205mm 100mm L W H Net weight ...

Page 14: ...m the date it sold out Damages not caused by misuse or vandalism all belong to warranty exclude the following case Balance out of warranty Damage caused by misuses such as long time under sunshine or receive corrosion The user does not obey the operation manual such as overloading the balance Without our authorized department or technician repairing the user disassembly the balance on his own Any ...

Page 15: ...ws the second calibration weights twinkling 4 Put on the weights on the pan accordingly it shows when it is stable it shows the weights remove the weights it turn to zero automatically the calibration is complete Note We suggest you do external calibration too when you begin to use the balance External calibration 1 Remove all objects from pan make sure the balance show zero 2 Press CAL key for 5 ...

Reviews: