4
【
In order to ensure the precision of weighing , please preheat the balance completely before do calibration
】
※
No load on the balance , balance already tared , inner weigh signal stable.
If the above conditions not fulfilled , it will display error message.
If the above conditions fulfilled , it will display calibrate the scale value you need.
Outside Calibration
button
(
command
)
steps panel display
short press[CAL] 1.Balance goes to Zero position
0.0000
g
long press [CAL] 2. About 5 seconds
CAL
Release
[CAL]
------
The scale value twinkles for 3 sec.
〖
200.0000
〗
3. Put on the indicated scale.
Display the calibrated scale value after 5 sec.
200.0000
g
4. Take off the scale, complete calibration.
0.0000
g
Application Function Operation
Choose the Mode:
button
(
command
)
steps panel display
Long press
[
MODE
]
Enter the choice for mode
Normal weigh mode
NORMAL
g
Count mode
COUNT
pcs
Percent weigh mode
PERCENT
%
Density measure mode*
DENSITY
d
*Density measure mode is an option, if you need please contact our company or authorized distributor.
Long press the button
[
MODE
]
, all the modes will display on the screen in a cycle, when the mode you want display on the screen , release the
button
[
MODE
]
to confirm the balance work mode.
Counting Function
Aim
By this procedure the user weigh the total weight of the certain samples which with similar single