Page 14
Standard Calibration
7. Allowable Conductivity standards are listed in section 1.3 and should
be selected according to your range of measurement.
8. Place the sensor into a sample of Conductivity standard, so that it is
immersed at least to the vent hole in the white plastic cover. The white
plastic cover MUST be in place for correct readings.
Glass sensors must be immersed to the hole in the glass body. See
the diagrams below.
DO NOT
place the sensor directly into the bottle of standard.
9. Discard the used sample of standard after use. It is advisable to use a
narrow sample vessel to minimise the use of standard solution.
10.When the reading has stabilised, Press the
key to calibrate. The
Aqua-CPA
will prompt the user to select the parameter if a
combination display is shown. Use the
or
keys to select
Conductivity.
The display will change to Conductivity if the
Aqua-CPA
is in TDS or
Salinity Mode during calibration.
The
∗∗∗∗
will now be replaced by a decimal point if calibration was
successful.
The
Aqua-CPA
is now calibrated for Conductivity.
The TDS and Salinity values are derived from the Conductivity reading
and do not require a separate calibration.
11.Discard the used samples of standard.