9
3. Specifications
IEC 1010 Over Voltage:
CAT II - 1000V
CAT III - 600V
Pollution Degree 2
Temperature for gauranteed accuracy: 23ºC +5ºC
DC VOLTS
Range
Res.
Accuracy
400mV
0.1mV
Impedance:
4V
0.001V
+/-(0.5% of r 2 digits)
10M
W
40V
0.01V
Overload Protection:
400V
0.1V
600VDC or AC RMS
600V
1V
(45Hz to 100Hz Frequency Response 400mV Range
)
AC VOLTS
(45Hz to 450Hz Frequency Response All Other Ranges)
Range
Res.
Accuracy
400mV
0.1mV
+/-(1.2% of r 5 digits)
Impedance:
4V
0.001V
+/-(1.2% of r 3 digits)
10M
W
40V
0.01V
Overload Protection:
400V
0.1V
+/-(1.5% of r 3 digits)
600VDC or AC RMS
600V
1V
8
DC Microamps
Range
Res.
Accuracy
40uA
0.01
m
A
+/-(0.8% of r 2 digits)
Overload Protection:
400uA
0.1
m
A
0.5A / 600V Fuse
e
*DC microamps are measured with the test leads in series with
the circuit under test.
OHM (
Resistance,
W)
Range
Res.
Accuracy
400
W
0.1
W
±(1% of r5 digits)
4k
W
0.001k
W
±(1% of r2 digits)
Overload Protection:
40k
W
0.01k
W
250VDC or AC RMS
400k
W
0.1k
W
4M
W
0.001M
W
±(3% of r5 digits)
40M
W
0.01M
W
±(3% of r10 digits)
Frequency (Hz)
Range
Res.
Accuracy
4.000kHz
0.001kHz
±(1% of r2 digits)
40.00kHz
0.01kHz
400.0kHz
0.1kHz
Overload Protection:
4.000MHz 0.001MHz
250VDC or AC RMS
40.00MHz 0.01MHz
400.0MHz 0.1MHz
UL 3111 Pending
(
45Hz to 100Hz Frequency Response 200A to 400A)
AC AMPS
(45Hz to 450Hz Frequency Response Below 200A)
Range
Res.
Accuracy
40A
0.01A
+/-(3.0% of r 5 digits)
400A
0.1A
e
Warning:
Use only correct type and overvoltage
category rated test leads. Remove the
temperature probe when using the test
leads or amp clamp.
Summary of Contents for 270
Page 1: ...270 Digital Clamp on Meter Instruction Manual ...
Page 37: ...37 mber mber 0RB ...
Page 38: ...38 ...
Page 39: ...39 ...