D. MEASUREMENT TECHNIQUES
1.
Controls and function:
Push Button
RANGE
Activates manual range
REC
Activates the Min/Max/Avg mode.
REL %
Activates REL% mode.
EDIT
Activates the EDIT mode for the compare
and relative % function.
Activates Back light for the
LCD(Automatically turns off after approx.
35 sec.)
COMP
Activates Compare mode.
HOLD
Activates tow-hold data hold mode
Rotary Switch
OFF
Turns the instrument off.
Selects the DC mV function
Selects the DCV function
Selects the ACV function
Selects the DCV function
(Push ORANGE button to activate ACV)
(Only192)
Selects the Diode test function
Selects the Ohm function (Push ORANGE
button to activate continuity buzzer.)
CAP
Selects the Capacitance function
14
3.
Specifications (cont.)
13
j. frequency
Range
Resolution
Accuracy
Overload protection
50Hz
0.001Hz
500Hz
0.01Hz
+/-0.05% of reading, 600V DC or AC
5Khz
0.0001Khz
+/- 5digits
peak
50Khz
0.001Khz
500Khz
0.01Khz
k. Adaptor
Model
Display
A301(Temp Adaptor C/F)
1mV/C, 1mV/F 0000.0 C/F
A254(Current Adaptor 10/60A)
100mV/A,
00.000A
10mV/A
000.00A
A256(Current Adaptor 40/400A)
1mV/A
0000.0A
A296(Current Adaptor 400/1000A) 1mV/A
0000.0A
l. General Specifications
Max. volt. between
1000V
any input and Ground
Fuse protection
mA :
0.5A/600VAC;
A :
10A/600VAC
[mA :
0.44A/1000VAC;
A :
11A/1000VAC(Only 194)
]
Display Type Digital : 50000 count, 4 times/S update Bar
graph : 51 segment
Operating temp.
0°C to 40 °C (32°F to 113°F)
Storage Temp.
-20°C to 60°C (-4°F to 140°F)
Relative Humidity
0% to 80% ( 0° ~ 35 °C/ 32° ~ 95°F)
0% to 70% ( 35° ~ 55 °C/ 95° ~ 131°F)
Power Supply
9Volt battery
Battery Life
Size(H x L x W)
Weight