TPC PC 2530 User Manual Download Page 2

 

Table of contents.  

 
1. Cover Page  
2. Table Of Co Markings 
3. Initial Setup  
4.  Initial Setup 
5. General use instructions 
6. Care and maintenance 
7. Troubleshooting  
8. Back Page Warranty Statement  
 
 
 
 
 

Markings 

Figures 

 

 

Notice:

 Risk of losing accessories and documents 

 
 

Caution:

 Risk of injury or malfunction 

 

 

Read carefully of the manual.  
 
 

 
 
Ground Protective 
 

 
 
Electron garbage bin  
 
Producing date 

 

On 
 
Off

 

 

 

Troubleshooting 

Unit will not power on:  

 

Check power cord.  
Check power switch.  
Check fuse.  Fuse is 250V 15Amp 
5X20mm 
If no green light is illuminated then 
there is no power to the main switch. 
Check fuse. Check outlet.  
Check internal PCB fuses on MAIN 
PCB. 

Suction is low  

Check suction Valve Spools and verify 
only one is in the open position.  
 
Check Solids Waste container. Clean 
or replace if needed.  
 
Check for internal leaks inside the 
machine.  
 

Unit is very loud when the  
motor is running.  

The motor should ramp un in rpms 
once in the on position.  Check for any 
suction clogs if the motor appears to 
be strained. If the motor makes loud 
screeching sounds, discontinue use 
and contact TPC support.  

The unit powers on and when 
I remove the suction valve 
the motor will not turn on.  
 

Check internal waste container and 
verify its empty. Check level probes 
inside the tank and verify they are 
clean. If the probes become corroded 
you may need to clean them with an 
abrasive cloth or pad.  

The fuse continues to blow 

Short or fault in the electrical circuit. 
Call TPC tech support for further 
assistance.  

Waste continues to drain out 
of the back of the machine 
when in use.  

 Check the internal waste container 
check valve. Clean and reseat if 
necessary.  

  

O                   

Reviews: