7
Light
head
adjustment
•
If
you
experience
a
downward
drift
with
your
operatory
light
head
please
follow
the
procedure
below.
•
Tighten
both
set
screws
on
the
back
side
of
the
operatory
light.
There
is
one
located
on
each
side
of
the
dental
light
head.
Reference
image
below.
To
tighten
the
set
screw
turn
it
clockwise.
Do
not
over
tighten
the
set
screw
as
damage
may
occur.
Vertical
Arm
adjustment
•
If
you
experience
a
downward
drift
with
your
operatory
lights
vertical
arm
please
follow
the
producer
below.
•
With
one
hand
lower
the
vertical
arm
till
you
see
the
adjustment.
See
image
below.
Adjust
the
inner
nut
till
the
desired
tension
is
achieved.
Less tension
More tension
Turn
tension
nut
counter
clockwise
if
vertical
light
arm
drifts
up.
Turn
tension
nut
clockwise
if
vertical
light
arm
drifts
down.