Summary of Contents for L700 Lucent Dental Light

Page 1: ...1 Advance Operatory Dental Light Operation and service manual TPC 851 S Lawson St City Of Industry CA 91748 P626 810 4337 Fax 626 810 4245 www tpcdental com REV 1 1 13...

Page 2: ...nts 5 Operation instructions 6 Light head adjustment 7 Vertical arm adjustment 7 Replacement of light bulb 8 Replacement Parts List 9 CLEANING DISINFECTING GUIDE 12 Warranty 13 TROUBLE SHOOTING GUIDE...

Page 3: ...matio of symb nating Curren ectrical shock er is disconne g this proced Failure to care bed procedu o the equipm e earth Grou n ols nt present Mak ected before ure efully follow re may result ment or...

Page 4: ...Insert the end of wire through the silver light bushing Place the end of the wire through the light post and insert the light bushing into the light post and press down to insure proper seating of th...

Page 5: ...5 Overview of components Light Post Transformer Housing Transformer lid Light bushing Horizontal Arm Light Knuckle Glass Reflector Deflector Power Switch Light Handle Vertical Arm...

Page 6: ...our hand on the light handle and move it to the desired location You can move the light head up and down and left and right When turning the light head left and right it will stop don t force any furt...

Page 7: ...urn it clockwise Do not over tighten the set screw as damage may occur Vertical Arm adjustment If you experience a downward drift with your operatory lights vertical arm please follow the producer bel...

Page 8: ...ushing it down then pulling it back towards the center of the light bulb socket Retaining spring Once the spring is removed then disconnect the light bulb from the wire Slide the high temperature heat...

Page 9: ...002 Transformer Housing 4 L700 0003 Horizontal Arm 5 L700 0004 Counter Sunk Pin 6 L700 0005 Swivel Arm Knuckle 7 L700 0006 Bridge ARM Connector 8 L700 0007 Set Screw 4 X 5 mm 9 L700 0008 Yoke Joint Co...

Page 10: ...10...

Page 11: ...10 L700 710 Flat Washer 11 L700 711 Wave Washer 12 L700 712 Nut For Joint Bolt 13 L700 713 Yoke Side Cover CAP 14 L700 714 Spring Holder For Light Bulb 15 L700 P107 Halogen Light Bulb 16 L700 716 Gro...

Page 12: ...finishes of dental equipment and are not recommended Use of these products will void your warranty Chemical Composition Trade Names Strong Phenols Lysol Lysol 2 Lysol Phenol Alcohol Professional Coe...

Page 13: ...ng from improper installation misuse or accidents incurred in shipping and handling All claims against the freight carrier must be initiated at the time the damaged items are received The claim is the...

Page 14: ...sformer Main power switch is bad Replace main power switch Light turns on and off then the light head is moved Short or open in power line form transformer Check each power line for short or open Then...

Reviews: