Installation requirements

Scan room

• Before installing the gantry, check the maximum permis-

sible floor load.

• The scanner emits radiation.  X-ray shielding must be

provided around the scan room and the entrance in
accordance with all local requirements and regulations.

• The ceiling should be at least 2,500 mm high to permit

the use of a contrast medium injector.

• Wiring pits and ducts are required for routing cables that

connect the various units.

Operator’s room

• An observation window is required for monitoring the

scan room.  X-ray shielding of the window glass must be
provided in accordance with all local requirements and
regulations, and the bottom of the window frame should
be 90 cm from the floor.

• Wiring pits and ducts are required for routing cables that

connect the various units.

• The operator’s room should have entrances for access to

the corridor and the scan room.

Cable connections between units

Checks before bringing-in the unit

• Check in advance the width of the corridor, the dimen-

sions of the entrance, and the dimensions and maximum
allowable load of the stairs and elevators to ensure that it
is possible to bring-in the unit safely and without difficulty.

• Minimum external dimensions of the entrance used for

bringing-in the unit are as follows:

Width:

980 mm

(38.6 in)

Height:

2,070 mm (81.5 in)

• The corners of corridors should be as illustrated below.
• Elevator minimum load: 2,000 kg (4,400 lb)

W

W

>

1,800 mm (70.9 in)

MPDCT0290EA

Breaker box

System

transformer*

Patient couch

Gantry

Console

(CPU cabinet 2)

Console

(CPU cabinet 1)

(Scan room)

(Operator's room)

Speaker

5

12

17

17

5

17

9

*: When line voltage is 200 V 

±

10 V, the system transformer is not required.

Unit: m

Summary of Contents for Activion 16

Page 1: ... data acquired using thin slices images with reduced partial volume effects can be obtained Outstanding operability Operability is improved as described below In 3D image processing or time consuming image pro cessing such as for regions in which calcified areas are superimposed on contrast medium bone elimination can be performed easily while observing reference images Most 3D images can be gener...

Page 2: ...r synchronization system CKIS 004A Orbital synchronized scan system CKOS 001A Pediatric scanning system CHKS 002A Not available in the U S A PERFORMANCE SPECIFICATIONS Scan parameters Scan regions Whole body including head Scan system 360 continuous rotate rotate Scan plan programming More than 360 different sequences can be pre pro grammed Scan time CT scan 0 48 s Partial 0 75 1 1 5 2 and 3 s Sca...

Page 3: ...image Real time helical reconstruction time 12 images s 0 083 s image 1 slice 512 512 matrix MPDCT0290EA Patient couch Vertical movement System Hydraulically driven Speed of vertical movement Up 16 to 24 mm s 50 Hz 19 to 28 mm s 60 Hz Down 20 to 30 mm s Stroke Approx 588 mm Minimum couch top height Approx 310 mm Maximum couch top height Approx 944 mm Couch top movement System Motor driven or manua...

Page 4: ...can rate Max 133 scans 100 s 0 75 s scan 133 rotations e c i l s g n i t t e S s s e n k c i h t h c t i p l a c i l e H 0 1 0 2 m m 5 0 m m 0 5 m m 0 0 1 m m 1 m m 0 0 1 m m 0 0 2 Image reconstruction Number of images Max 4 images scan Image interval Reconstruction is possible in increments of 0 1 s Reconstruction time Min 0 5 s Real time reconstruction time 12 images s 0 083 s image 1 slice 512 ...

Page 5: ...unction Software control function key Multi frame display Reduction cut off display ROI processing Inset scanogram display Selective related information display Cine display Image display speed Variable Scanogram CT image switching Show scano line zoom hide scano line Slice feed playback CineView High speed image feeding using the mouse or keyboard Image processing Scanogram processing Slice posit...

Page 6: ...U TIFF conversion Filming Ethernet Toshiba protocol DICOM PRINT SCU Sheet editing function using virtual film T mode Related information items such as the patient name are dis played in the footer area using a larger font Note To use T mode the laser imager must support 2048 pixels x 2404 pixels for a 1 x 1 frame Auto filming in eXam Plans IMAGE QUALITY Noise Standard deviation Less than 0 5 Scan ...

Page 7: ...built into the gantry and the system employs a high frequency inverter for generat ing and stabilizing the high voltage supplied to the X ray tube The generator includes electronic circuits for control ling the speed of the rotating anode in the X ray tube Use of a high frequency inverter system results in high power output combined with excellent stability In addition the system is compact and li...

Page 8: ...for rapid patient scanning during a single breath hold and for high defini tion three dimensional and MPR imaging Real time helical reconstruction mode makes it possible to observe the images being scanned in real time at a maximum at 12 frames per second This mode shows any shift in the slice position in real time and helps the operator to check the scan field on the image the con trast study tim...

Page 9: ...er scan Real time reconstruction is possible in scanoscopy CT images can be reconstructed in 0 5 second for 0 75 second routine scans Ease of operation is ensured by incorporating a hybrid keyboard mouse driven menus and a large color LCD screen The couch top can be lowered very near the floor simpli fying patient transfer COMPLIANCE Council Directive 93 42 EEC Concerning Medical Devices Medical D...

Page 10: ...nsation 1 When scanning is not performed 2 When scanning is performed continuously at the maximum rated output of the system Minimum area for installation For the long patient couch version 22 m2 For the short patient couch version 21 m2 NFB 150 A NFB 100 A In case of 3 phase 200 V In case of 3 phase 400 V System System transformer option Ground resistance As per applicable legal requirements Grou...

Page 11: ...quired for routing cables that connect the various units The operator s room should have entrances for access to the corridor and the scan room Cable connections between units Checks before bringing in the unit Check in advance the width of the corridor the dimen sions of the entrance and the dimensions and maximum allowable load of the stairs and elevators to ensure that it is possible to bring i...

Page 12: ...12 OUTLINE DRAWINGS Gantry and Patient Couch ...

Page 13: ...MPDCT0290EA 13 OUTLINE DRAWINGS 344 13 5 556 21 9 1 910 75 2 900 35 4 2 070 81 5 TILT ANGLE Gantry ...

Page 14: ...OUTLINE DRAWINGS 1 862 73 3 2 690 105 9 588 23 1 1 038 40 9 STROKE 450 17 7 2 187 86 1 2 190 86 2 470 18 5 630 24 8 430 When the arm up holder is mounted 17 Patient Couch for the long patient couch version ...

Page 15: ...EA 15 OUTLINE DRAWINGS 1 562 61 5 2 390 94 1 588 23 1 1 038 40 9 STROKE 450 17 7 1 887 74 3 1 890 74 4 470 18 5 630 24 8 17 430 When the arm up holder is mounted Patient Couch for the short patient couch version ...

Page 16: ...of Toshiba Medical Systems Corporation This document may include trademarks and registered trademarks of other companies Model TSX 031A 2007 04 TME KI Toshiba Medical Systems Corporation Nasu Works meets the Environmental Management System standard ISO 14001 Toshiba Medical Systems Corporation meets internationally recognized standards for Quality Management System ISO 9001 ISO 13485 Printed in Ja...

Reviews: