Programming Example 1:
No Wait-Until, No Auto-Off
Figure A1:
Global Wait-Until and Auto-off are Both Disabled
If the “wait until” and “Auto-Off” features are set to “no”, then Figure A1 above illustrates
a three step program where the first step has a set point of 37C and the countdown
timer starts at 30mins. When the first step is running, the plate will heat until the plate
temperature is 37C. The plate temperature will then remain at 37C for the remainder of
the 30minutes. When the timer reaches 00:00:00, step 2 loads its values and begins to
execute. In step 2, the set point is 95C and the countdown timer value is 1 hour. In
step 3 the set point is -4C for 30mins. When the program ends, the timer will begin
counting up until the user presses a button on the IC25/35 front panel to return to the
default screen. In this example, the unit continues operation at the same set point as
the last programmed step.