4.1.4 A0/A1/A2/A3 Command
Change the measurement mode in the instrument.
This setting is stored even after the power is OFF. If “ST” command is sent without sending this
command, the measurement mode, which has been set before sending, is used for measurement.
Command
Measurement mode
A0
: AUTO
A1_#####
: FREQ
“_” is space. “#####” is frequency.
0
–
250 Hz
A2_#####
: MANU
“_” is space. “#####” is integral time.
A3
: SYNC
20
–
15000 ms
4.1.5 KW [n] Command
Write the correction factor in the inst
rument. Input 0 to 430 to “n”. It is possible to be applied to 600
– 1030nm at intervals of 1nm. The factor can be input by one waveform and is sent with the form
“KW [n]_#. ###”.
When the instrument receives the correction factor, it returns “OK” as the reception check command.
Example) KW0_1.000+(Cr
・
Lf)
“_” is space.
Input the correction factor 1.000 to “n=0 (600nm)”.
KW0
: 600 nm
KW1
: 601 nm
KW2
: 602 nm
• • • • •
• • • • •
KW429 : 1029 nm
KW430 : 1030 nm
“A0”, “A1”, “A2” or “A3”+(Cr·Lf)
“OK”+(Cr
・
Lf)
PC
Command/Communication direction
SR-NIR
“KWn_#.###”+(Cr
・
Lf)
“OK”+(Cr
・
Lf)
“OK”+(Cr
・
Lf)
“KWn_#.###”+(Cr
・
Lf)
PC
Command/Communication direction
SR-NIR