3.11 Using a Correction Factor
Select whether to use a correction factor.
A correction factor is used to correct the measurement values. (A measurement value is multiplied by
the correction factor.) There are two types of correction factors as described below.
Factor (1 nm):
Corrects one spectral data per 1 nm. There is one correction factor group.
Set ON or OFF to use or not to use the correction factor.
Factor (XYZ):
Corrects a tristimulus XYZ value. There is one correction factor group. Set
ON or OFF to use or not to use the correction factor.
Use the following procedure to set the correction factors:
1
Switch to the function mode to display the [FACTOR]-[Factor ON/OFF].
☞
'3.1.2 Entering/Returning from the Function Mode'
2
Press the [CHANGE] switch.
The right portion of the “Factor (1 nm)” will blink.
3
Press the [ROTATION] switch to select the desired value.
Each time the switch is pressed, the value changes between ON and OFF.
To skip the “Factor (1 nm)” without changing it, proceed to step 4 without pressing the
[ROTATION] switch.
4
Press the [ENTER] switch to accept the setting for the factor (1 nm).
If the [ENTER] switch
is pressed after selecting the setting for the “Factor (XYZ)”, the
next setting item screen will be displayed.
Correction Factor Input Screen
☞
’3.12 Displaying and Editing the Correction Factor’