41
GENERAL INFORMATION ON USAGE AND MAINTENANCE
ELECTROMAGNETIC COMPATIBILITY
This product conforms to the EMC standard (IEC 60601-1-2 Ed.3: 2007).
a) MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in the ACCOMPANYING
DOCUMENTS.
b) Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
c) The use of ACCESSORIES, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replacement
parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the
EQUIPMENT or SYSTEM.
d) The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equipment. If adja-
cent or stacked use is necessary, the EQUIPMENT or SYSTEM should be observed to verify normal
operation in the configuration in which it will be used.
e) The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other than those
specified may result in increased EMISSION or decreased IMMUNITY of the EQUIPMENT or SYS-
TEM.
Item
Cable
shield
Ferrite
core
Length(m)
ACCESSORIES
DIGITAL CAMERA UNIT DC-4
-
-
-
BACKGROUND ILLUMINATION BG-2GN
-
-
-
CABLES
AC Power cord (for AC adapter)
Not Used
Not Used
3.0
DC-4 TRIGGER cable
Used
Used
1.0
LAN cable
Used
Used
3.0
ILLUMINATION cable
Used
Not Used
1.6
AC adapter cable
Not Used
Used
0.15
Guidance and manufacturer's declaration - electromagnetic emissions
The SL-D301 is intended for use in the electromagnetic environment specified below. The customer or
the user of the SL-D301 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The SL-D301 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The SL-D301 is suitable for use in all establishments includ-
ing domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies