SPECIFICATIONS

Guidance and manufacturer's declaration - electromagnetic immunity

The SL-1E is intended for use in the electromagnetic environment specified below.

The customer or the user of the SL-1E should assure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance

 level

Electromagnetic environment - 

guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150kHz to 80MHz

3 V/m

80MHz to 2.5GHz

3 V

3 V/m

Portable and mobile RF communica-

tions equipment should be used no

closer to any part of the SL-1E, includ-

ing cables, than the recommended

separation distance calculated from the

equation applicable to the frequency of

the transmitter.

Recommended separation distance

d

 = 1.2 

d

 = 1.2 

   80MHz to 800MHz

d

 = 2.3 

   800MHz to 2.5GHz

where 

P

 is the maximum output power

rating of the transmitter in watts (W)

according to the transmitter manufac-

turer and 

d

 is the recommended sepa-

ration distance in meters (m).

Field strengths from fixed RF transmit-

ters, as determined by an electromag-

netic site survey,

 a

 should be less than

the compliance level in each frequency

range.

 b

Interference may occur in the vicinity of

equipment marked with the following

symbol:

NOTE 1

NOTE 2

At 80 MHz and 800 MHz, the higher frequency range applies.

These guidelines may not apply in all situations. Electromagnetic propagation is

affected by absorption and reflection from structures, objects and people.

a

b

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV

broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic

environment due to fixed RF transmitters, an electromagnetic site survey should be con-

sidered. If the measured field strength in the location in which the SL-1E is used exceeds

the applicable RF compliance level above, the SL-1E should be observed to verify normal

operation. If abnormal performance is observed, additional measures may be necessary,

such as reorienting or relocating the SL-1E.

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

P

P

P

Summary of Contents for SL-1E

Page 1: ...INSTRUCTION MANUAL SLIT LAMP SL 1E...

Page 2: ......

Page 3: ...the eye open To understand and follow instructions when undergoing an examination WARNING To avoid injury to the patient s eye and nose pay particular attention while operating the instrument body Th...

Page 4: ...ng the base unit Disposal When disposing the instrument and or parts follow the local regulations for disposal and recy cling STORAGE PLACE USAGE PERIOD 1 When storing the instrument ensure that the f...

Page 5: ...ury CAUTION Ignoring or disregarding this display may lead to personal injury or physical damage Injury refers to cuts bruises sprains fractures burns electric shocks etc Physical damage refers to dam...

Page 6: ...Ask your dealer for service CAUTION Icon Prevention item Page To avoid electric shock do not touch the external connection ter minal and the patient at the same time 31 34 This instrument has been tes...

Page 7: ...e page 30 DISCLAIMERS TOPCON is not responsible for damage due to fire earthquakes actions or inactions of third persons or other accidents or damage due to negligence and misuse by the user and any u...

Page 8: ...Keep a safe distance from the patient s face to prevent injuries to the nose and eyes Gardez une distance de travail appropri e afin d viter le contact avec les yeux et le nez du patient CAUTION To p...

Page 9: ...ER COMPENSATION AND INTERPUPILLARY DISTANCE ADJUSTMENT 16 PATIENT POSITION 17 BASE OPERATION 17 OPERATION OF THE ILLUMINATION UNIT 18 TROUBLESHOOTING 19 TROUBLESHOOTING GUIDE 19 SPECIFICATIONS 20 ENVI...

Page 10: ...8 CONTENTS HRUBY LENS 35 10 MEASURING EYEPIECE 36 16 EYEPIECE 36 ACCESSORY DRAWER 37 APPLANATION TONOMETER 37 AUTOMATIC INSTRUMENT TABLE AIT 15 37...

Page 11: ...tch Power switch Pilot lamp Click stop Lamphouse cover clamping screw Lamphouse cover Filter selection lever Aperture and slit length control wheel Slit width control knob Slit rotation control ring M...

Page 12: ...eses are the quantities Some accessories are not included as standard depends on the model Chinrest paper 1 Focusing test rod 1 Spare illumination lamp 1 Spare chinrest paper pin 2 Spare fuse 1 Brush...

Page 13: ...type table w power supply 2 Unit type table w power supply 3 Rail cover 4 Chinrest unit 5 Power cable Article name Q ty 1 Main unit 1 2 Instrument type table w power supply 1 2 Unit type table w powe...

Page 14: ...A and Canada for U S A and Canada F1 F2 T1AL 250V 100 120V F1 F2 T1AL 250V 120V T500mAL 250V 220 240V T500mAL 250V 220V F3 F4AL 125V F3 F4AL 125V 2 Mounting the Table a To attach the table on the inst...

Page 15: ...e relocated to the right side of the unit top to prevent interference with the arm In this instance remove the four wood screws which attach the power supply to the unit top re position the power supp...

Page 16: ...e between the rail and table Remove the protective tape from the illumination unit and the hole cap while being careful to avoid touching the lens surface 5 Connecting Cables Connect the chinrest cabl...

Page 17: ...e sure that the instrument is always properly grounded 2 Illumination and Instrument Functions Turn the power supply on and observe that the illumination is passing through the opened slit controls Ch...

Page 18: ...er switch ON Set the brightness control switch at the 0 1 position Turn aperture and slit length control wheel until slit diameter becomes 14mm then turn slit width control knob to set slit width to m...

Page 19: ...nrest and forehead against the forehead rest Adjust the chinrest adjuster so that the patient s outer canthus is at the approximate height of the canthus marker BASE OPERATION 1 Horizontal gross adjus...

Page 20: ...PERATION OF THE ILLUMINATION UNIT 1 Change of slit size By operating the slit width control knob or turning the aperture and slit length control wheel ideal slit width slit length and spot size can be...

Page 21: ...till OFF Turn power switch ON The lamp socket is damaged from the heat Replace the socket P 31 The lamp has burned out Replace the lamp P 31 Fuse has blown Replace the fuse P 30 Slit light is too dim...

Page 22: ...Aperture diameters 14mm 10mm 8mm 5mm 3mm 2mm 1mm Slit angle Continuously rotatable from vertical to horizontal Filters Blue Red free Lamp 6V 20W halogen BASE Longitudinal movement 80mm Lateral moveme...

Page 23: ...1 2 Ed3 0 2007 EN60601 1 2 2007 AC 2010 UL 60601 1 2003 ISO 10939 1998 ISO 15004 1 2006 ISO 15004 2 2007 EC Medical Device Directive 93 42 EEC Weight W Table W Unit Table W O Table W O Table and Chinr...

Page 24: ...40 C Humidity 10 95 without dew condensation Air pressure 700hPa 1060hPa THIS INSTRUMENT DOES NOT MEET THE TEMPERATURE REQUIREMENTS OF ISO 15004 1 FOR STORAGE DO NOT STORE THIS INSTRUMENT IN CONDITION...

Page 25: ...adjacent or stacked use is necessary the EQUIPMENT or SYSTEM should be observed to verify normal operation in the configuration in which it will be used Guidance and manufacturer s declaration electr...

Page 26: ...o line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and Voltage...

Page 27: ...it ters as determined by an electromag netic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the follow...

Page 28: ...ation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 80MHz to 800MHz d 1 2 800MHz to 2 5GHz d 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12...

Page 29: ...expo sure limit If the value of radiance were reduced to 40 mW cm2 sr twice that time i e 6 min would be needed to reach the recommended limit The recommended exposure is based on cal culations arisin...

Page 30: ...h one comes into contact do not become conductive if the basic insulation fails The mode of operation continuous operation equipment Degree of protection against ingress of water IP 0 SL 1E is the ord...

Page 31: ...e A Hospital Grade Canada 120V 60Hz Type A Hospital Grade Symbol IEC ISO Publication Description Description French IEC 60417 5032 Alternating Current Courant alternatif IEC 60417 5008 Off power disco...

Page 32: ...The fuse will come out Replace it with a new fuse and then tighten the center of the fuse holder Always use the same type of fuse as indicated in the holder F1 F2 T1AL 250V 100 120V T500mAL 250V 220...

Page 33: ...ert the new lamp into the socket Place the assembly into the housing by aligning the guide with the groove in the lamp flange Note If the lamp and or socket is incorrectly attached uneven or partially...

Page 34: ...e base rail and shaft If the gliding plate or cross slide rail and shaft are dirty an unsmooth vertical or horizontal movement of the cross slide results Clean them by using a dry cloth 3 Cleaning the...

Page 35: ...ing Adjusting the diopter and pupillary distance Focus of slit image The base unit must move smoothly The components must be fitted in place correctly The chinrest unit must be fitted to the table uni...

Page 36: ...of the chinrest Insert the fixation target into either of the two mounting holes and then remove the cap to insert the fixation target plug into the chinrest jack Specifications Annular Fixation Targ...

Page 37: ...rm so that the groove of guide plate fits with guide pin on the microscope arm 3 Insert the lower end of the Hruby Lens shaft in the groove on the guide plate 4 Center the illumination and microscope...

Page 38: ...remely helpful when fitting toric contact lenses Specifications Scales Linear Scale 16mm minimum 0 5mm increments to be used at 10 magnification Angle Scale 360 minimum 5 readings Diopter compensation...

Page 39: ...ONOMETER Haag Streit AG Model R900 can be used AUTOMATIC INSTRUMENT TABLE AIT 15 Specifications Dimensions 510 W 450 D mm Table height 600 820mm Table size 490 500mm Weight approx 23kg Power consumpti...

Page 40: ......

Page 41: ...plate on the left of the base Period of Usage i e the purchase date Description of Problem as detailed as possible SLIT LAMP SL 1E INSTRUCTION MANUAL The 2011 version 2011 06 100TH Date of issue 3 Ju...

Page 42: ...SLIT LAMP SL 1E 44615 95997 Printed in Japan 1106 100TH 7...

Reviews: