95
SPECIFICATIONS
ELECTRIC RATING
Power supply voltage : AC100V~120V, 220~240V, 50/60Hz
Power supply input
: 280VA
CLASSIFICATION
•
The type of protection against electric shock: Class
I
equipment
This instrument is classified as Class
I
. Class
I
equipment does not depend only on basic insulation
against electric shock. It contains a mechanism that connects with the protective grounding system
of the facilities, so that the metal parts with which you come into contact cannot electrify you if the
basic insulation malfunctions.
•
The degree of protection against water damage: IPX0 (Main body), IPX8 (Foot switch)
•
The mode of operation: Continuous operation
Recommended separation distance between
portable and mobile RF communications equipment and the OMS-800
The OMS-800 is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the OMS-800 can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the OMS-800 as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150kHz to 80MHz
d = 1.2
80MHz to 800MHz
d = 1.2
800MHz to 2,5GHz
d = 2.3
0, 01
0, 12
0, 12
0, 23
0, 1
0, 38
0, 38
0, 73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1
NOTE 2
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
P