Collimate the staff A and push the [Meas] key. (step3)
7
Collimate the staff B and push the [Meas] key. (step4)
8
The complemented value (electrical reading pattern) is displayed.
Press the [D/R] key to set displayed electrical adjustment value, then
the software indicates the rod value that should be complemented for
the reticle adjustment.
Here the electrical adjustment is completed, however, we recom-
mend to check the optical reading (Reticle adjustment) with the fol-
lowing procedure always when the electrical adjustment is
performed.
9
Reverse the staff B to be optical reading pattern side. Take the reticle
cover off. Adjust the reticle adjustment screws by using the hexagon
wrench till the reticle is complemented height. In case that the reticle
adjustment is ensured prior to above electrical adjustment, you can
skip this step
9
and turn off the instrument.
10
Turn off the power supply.
+3.5mm
1.234m
Reticle adjustment screw
Reticle