58
MAINTENANCE
CLEANING THE COMPONENTS THAT COME INTO CONTACT WITH THE PATIENT
When the forehead rest and chinrest become stained, use a neutral tableware detergent and warm
water. Dip a soft cloth in the solution, squeeze out the excess water and then wipe off the stain.
DAILY MAINTENANCE
• For this instrument, dust may cause errors. When not in use, replace the nozzle cap and dust
cover.
• When not in use, turn off the POWER switch.
ORDERING CONSUMABLE ITEMS
• When ordering consumable items, tell the product name, product code and quantity to your dealer
or TOPCON at the address listed on the back cover.
USER MAINTENANCE ITEM
Product name
Product code
Product name
Product code
Chinrest tissue
40310 4082
Applicator
41601 8606
Monitor cleaner
44800 1001
Printer paper
44800 4001
Dust cover
42360 9002
Item
Inspection time
Contents
Inspection
Before using
• The instrument works properly.
• The objective lens must be free of stains and/or
flaws.
• Confirm whether the foreign object is attached
to the measuring nozzle and the area around
the measuring nozzle.
• Air check
• Confirm that the safety stopper setting and
measuring nozzle do not move to the patient's
side beyond the safety stopper setting position.
Cleaning
When the part is stained
• Objective lens
• External cover, control panel, etc.
Replacement
As required
• Printer form