CA-200F Corneal Analyser - Rev. 7 16/01/2012
5
PRECAUTIONS
This electronic instrument is a precision unit. Use and store it in a suitable place in normal temperature,
humidity and atmospheric pressure conditions and avoid exposure to direct sunlight.
To ensure proper functioning, install the instrument in a place not subject to vibrations.
Connect all the cables correctly before use.
Use the recommended mains voltage.
When the unit is not used, disconnect the power supply and protect it against the sun and dust.
In order to obtain accurate and reliable measurements, keep the measuring cone clean and dust-free.
This product is in compliance with the EMC standards (IEC 60601-1-2:2001).
-
ELECTROMEDICAL DEVICES require particular precautions for electromagnetic compatibility and
must be installed and set up based on the EMC information provided in the accompanying
documents.
-
Portable RF communication instruments may interfere with medical devices.
-
Using accessories and cables different from those provided with the instrument, except for the
cables sold by the equipment manufacturer as spare parts, may result in increased emissions and
reduce the immunity of the device or system.
-
The device must not be used in contact with other equipment.
If it is inevitable to use the device in contact with other instruments, check proper functioning in the
required configuration.
EMC table
Emission aspects
The CA-200F device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should ensure that it is used in the said environment.
Emission test
Compliance
Electromagnetic environment
– guidance
RF emissions
CISPR 11
Group 1
The CA-200F device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The CA-200F device is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3:2
Class A
Compliant
The device can be used in all buildings, including domestic buildings
and those directly connected to the public low-voltage power supply
network, that supplies buildings used for domestic purposes.
Voltage fluctuations/
flicker emissions
IEC 61000-3:3
Compliant