41
4.1.1 STRn COMMAND
Associated mode: Luminance / Illuminance / UV mode
When the BM-100 receive this command, the BM-100 returns measured data in a text format.
n: 0 (Auto range)
n: 1
– 4 (Manual range, Luminance / UV mode)
n
:
1 - 6 (Manual range, Illuminance mode)
PC
Command
BM-100
STRn+(Cr
・
Lf)
OK+(Cr
・
Lf)
Measured value
Data+(Cr
・
Lf)
END+(Cr
・
Lf)
If message “NG” returned
☞
‘
5.1 INSTRUMENT ERROR CODE
’
4.1.2 CAL COMMAND
Associated mode: Luminance / Illuminance / UV mode
When the BM-100 receive this command, the BM-100 starts zero adjustment.
PC
Command
BM-100
CAL+(Cr
・
Lf)
OK+(Cr
・
Lf)
Finish zero adjustment
END+(Cr
・
Lf)
Summary of Contents for BM-100 Series
Page 1: ...INSTRUCTION MANUAL Phosphorescence luminance meter Rev 8...
Page 2: ......
Page 54: ...52 OUTLINE DIMENSION BM 100 unit mm...
Page 58: ......