25
Care and check
6
Match the protruded part of the lamp attaching area with the lamp flange notch and
tighten them with the lamp attaching screw to secure the lamp
7
Place the lamp cover on.
8
Turn the power switch 'on'.
9
Make a trial projection to check that there is no illumination irregularity. If there is any
irregularity, turn the power off again and check that the lamp is installed properly.
Replacing the Fuse
1
Turn the power switch 'off' and disconnect the power supply plug.
2
Simultaneously squeeze both sides of the fuse holder and then remove the fuse
holder. The fuses will be removed together with the holder.
3
Remove the blown fuses and insert the spare fuses on the cap of the fuse holder
and install it as before (Press to install it). Use the attached fuses or the fuses spec-
ified below.
Warning
Unplug the power cable before removing the fuse cover to replace
the fuse. Removing the fuse cover with the power cable plugged
in can cause electric shock. Do not install the power cord on the
body with the fuse cover removed.
Warning
Use only attached fuses (For 100, 120V range: T-1.6A, 250V For
220, 230, 240V range: T-1A, 250V). Using other fuses may cause
a fire in the event that the instrument fails.