11
TOPCOM BPM Wrist 7500
ENGLISH
To erase all data from one memory zone:
•
Press the
START/STOP
button to activate the display.
•
Press the
UP
or
DOWN
button to select the desired
memory zone.
•
Press the
MEMORY
button to see the average of all
measurement results.
•
Press the
MODE
and the
DOWN
button simultaneously to
erase all data from the selected memory zone.
15 Technical specifications
*Specifications are subject to change without notice.
Power source
Measurement method
Measurement zone
Accuracy
Pressure sensor
Inflation
Deflation
Memory capacity
Auto-shut-off
Operation environment
Storage environment
Weight
Wrist circumference
Limited patient
Dimensions
K.B.
Classification
3V DC: Two LR03 (AAA) Non rechargeable batteries
Oscillometric
Pressure: 0 ~ 300 mmHg
Pulse: 40 ~ 199 beats/minute
Pressure: ± 3 mmHg
Pulse: within ± 5% of reading
Semi-conductor
Pump driven
Automatic Pressure release valve
3 Memory zones for each 40 memories
1 minute after last key operation
+10°C ~ + 40°C
85% RH
-20°C ~ + 70°C
85% RH
148g (including batteries)
135 ~ 195 mm
Over 18 years old
74(L) x 84(W) x 32(H) mm
Short time operation 2 minutes
• Internally powered equipment
• Classification of type BF (applied part)
• IPX0
• Not suitable for use in presence of flammable anaes-
thetic mixture with air or with oxygen or nitrous oxide
• Continuous operation with short-time loading
≤
≤
The CE symbol indicates that the unit complies with the
essential requirements of the directive 93/42/EEC.