Top Tronic T1151 Instruction Manual Download Page 7

The Multifunction Tester has the features of the Insulation Tester plus the 
MOV and  GAS Arrester Test. 
To access the MOV test,                You need the depress the 1000V and 
500V keys                                       Simultaneously And To access the GAS 
Arrester Test, you                            need to depress the 250V and Ohm Key 
simultaneously.

The Multifunction Tester has 

everything                                                              

You'll ever Want from an 

electrical                                                                  

Test and Measuring tool.
Before any test is                                                             performed, (provided 
the test leads are                                                         connected properly and 
the fuse is intact) it                                                     make a voltage test to 
assert  that there is no                                             voltage on the device or 
circuit under test. 
If there is a voltage which could be a problem for the test, the tester, 
automatically, switch to the Voltmeter and shows the voltage on the display. 
Be it AC or DC.
If voltage is present on the leads before test is started, the selected test is 
cancelled and the keypad is disabled, preventing errors of operation. This 
makes this tester one of the safest to operate to date.
Once any voltage is cleared, test can begin. Should you want to measure 
Insulation Resistance, you can select by using the 250V, 500V and the 1kV 
test Voltage which all go into the Giga-Ohm range. Should you want to test 
continuity, use the Low ohms scale, down to 0.01Ohm and also the buzzer. 
You can null the fuse and test leads using the Auto-Null feature. Of Course 
on this model, Auto-hold is featured. You will appreciate the superior Safety 
of Auto-Hold when you decide to visually look after your arms and fingers  
without needing to check the display. Auto-Hold is always turned ON and 
with it, you can simply touch the voltage and remove the test probes. The 
last valid voltage will be on the display when you want to see it. 
While Dangerous voltages are present on the leads, it beep too, signalling 
the user that he is touching dangerous voltage.

-5-

Summary of Contents for T1151

Page 1: ...INSTRUCTION MANUAL MULTIFUNCTION INSULATION CONTINUITY VOLTAGE TESTER Model T1151...

Page 2: ...r 05 Notes 06 Features 07 09 Pre testing Safety 10 Operating Instructions 11 13 Inside Lid Instructions for Insulation Tester 14 Inside Lid Instructions for Multifunction Tester 15 Very Important Auto...

Page 3: ...rsonnel vulnerable to exposure caused by faults at or fed from the system under test transferred potentials from remote test grounds and inadvertent line energisations While the probability of the occ...

Page 4: ...fuse use only the type specified HBC fuse and insert correctly into the fuse holder Always double check the lead connections before making any measurements For increased safety use fused test leads o...

Page 5: ...and electrical installations are protected by MOVs and GAS arresters The Multifunction Tester can test these devices to establish if the devices are still operating correctly or not Energy conservatio...

Page 6: ...disabled To disable EnerSave the user need while starting a test to depress the TEST key for more than 3 Seconds After 3 Seconds a short beep will be heard meaning that EnerSave has been disabled Once...

Page 7: ...the selected test is cancelled and the keypad is disabled preventing errors of operation This makes this tester one of the safest to operate to date Once any voltage is cleared test can begin Should y...

Page 8: ...Notes 6...

Page 9: ...isplay Voltmeter There is no Voltmeter Key as this is the default mode of this instrument The tester switch to the voltmeter mode after start up This is an automatic AC DC voltmeter This mode is the d...

Page 10: ...unning much longer then depress and keep depressing the TEST key for more than 3 Seconds the tester will beep when EnerSave is disabled For longer tests duration every time you start a test you need t...

Page 11: ...as well as the equivalent Max AC voltage which could be utilized on this device GAS Arrester Selection keys On the Multifunction Tester The GAS test can be accessed by depressing the 500V and Ohm key...

Page 12: ...all the safety features will only work if you have proper connections prior during and post testing to the circuit under Test All safety features can only work if the fuse is intact and correct Follow...

Page 13: ...measured and displayed Note that battery voltage is always monitored during use of the instrument Should the battery become too low the low battery symbol will be displayed Ac and Dc voltages are aut...

Page 14: ...a Smart Program which TM EnerSave Save Energy when ever possible by limiting the Test Duration PI Polarisation PI this is the ratio of the Insulation Index Resistance at 10 Min divided per the Insulat...

Page 15: ...d by MOVs It is now easy to test these devices to ensure their proper working and replace them if found damaged The knee Voltage is shown on the display Pressing the 250V and W keys simultaneously wil...

Page 16: ...uto Null Leads and Fuse resistance The auto null value is saved in memory Voltmeter with Automatic Hold Auto Hold helps the user taking measurements without looking at the display The user can focus i...

Page 17: ...y Test 0 01W to 1999W Press again to Auto Null Leads and Fuse resistance The auto null value is saved in memory Voltmeter with Automatic Hold Auto Hold helps the user taking measurements without looki...

Page 18: ...leads make contact at any time before during and after the test Once a test is finished the testers will automatically discharge capacitive or inductive circuit of their charge The discharge can be ob...

Page 19: ...mA Typical Continuity Open Circuit Voltage of 5V dc Nominal Voltage 1mA on all Insulation Ranges Buzzer ON OFF by Key Always ON Always ON Leads Auto Null key Test Auto Stop Display Customization for O...

Page 20: ...dc AUTOMATIC BATTERY TEST When the tester starts it test it s batteries by drawing a heavy current from the batteries During that heavy current it measures the battery voltage and display it for a few...

Page 21: ...nus one digit Below 1 ohms it s better than 5 of readings plus minus 3digits Above 200 ohms up to 1999 ohms it s better than 3 of readings plus minus 3 digits The Continuity has an auto nulling facili...

Page 22: ...ted That mean that there have no potentiometers or moving part in this product All the correction factors are saved into the internal memory BATTERIES The testers can work with 8 x 1 5V batteries The...

Page 23: ...od condition Cleaning Use a damp cloth to clean the case Do not use chemicals REPLACE BATTERIES This instrument operate well with Alkaline 1 5V or rechargeable 1 2V batteries It could be operated with...

Page 24: ...to 700Vac Resolution on both voltage types 1V Accuracy on both voltage types 1 5 of FS INSULATION TEST Resistance Range 250V Test Voltage 2k 2G Ohms 500V Test Voltage 4k 4G Ohms 1000V Test Voltage 8k...

Page 25: ...cy 300 to 1999 Ohms 2 of Rdg Continuity Resolution 2 Counts AUTO NULL Auto Null Value Saved in no volatile Memory Auto Null Threshold 5 Ohms BUZZER Buzzer Threshold 3 Ohms MOV TEST Test Voltage 5 1020...

Page 26: ...erial Polycarbonate A BS Weight Approx 655g battery included Display Liquid Crystal Display ENVIRONMENTAL Operating temperature Range 1 C to 55 C not in full sun Storage Temperature 20 C to 70 C CLEAN...

Page 27: ...any way so as in our sole judgement to injure their stability or reliability or which have been subject to misuse abuse misapplication negligence or accident or which have had the serial numbers alte...