108
Capacitance(c) and Dissipation (D)
Range
Display Range
Accuracy
Ce
Accuracy
De*
Recommended
Equivalent
Mode
10
0Hz/12
0Hz
20mF 4.000mF
20.000mF
5.00%+5
digits
0.0500
Series
4mF 400.0
μ
F
3.9999mF
1.00%+3
digits
0.0100
Series
400
μ
F 40.00
μ
F
399.99
μ
F
0.35%+2
digits
0.0035
Series
40
μ
F 4.000
μ
F
39.999
μ
F
0.10%+2
digits
0.0010
Series
4
μ
F 400.0nF
3.9999
μ
F
0.10%+2
digits
0.0010
----
400nF 40.00nF
399.99nF
0.10%+2
digits
0.0010
Parallel
40nF 4.000nF
39.999nF
0.35%+3
digits
0.0035
Parallel
4nF 0pF
3.999nF
1.25%+5
digits
------
Parallel
1k
Hz
1000
μ
F 400.0
μ
F
999.99
μ
F
2.00%+5
digits
0.0200
Series
400
μ
F 40.00
μ
F
399.99
μ
F
1.00%+3
digits
0.0100
Series