3. CLEANING AND DISINFECTION INSTRUCTIONS
Before any cleaning of the instrument or the base unit, ensure the power
lead is disconnected.
Only manual non-immersion cleaning as described should be used for this instrument. Do not
autoclave or immerse in cleaning fl uids. Always disconnect power supply from source before
cleaning.
1. Wipe the external surface with a clean absorbent, non-shedding cloth dampened with
de-ionised water / detergent solution (2% detergent by volume) or water / isopropyl
alcohol solution (70% IPA by volume). Avoid optical surfaces.
2. Ensure that excess solution does not enter the instrument. Use caution to ensure cloth is
not saturated with solution.
3. Surfaces must be carefully hand-dried using a clean non-shedding cloth.
4. Safely dispose of used cleaning materials.
4. INSTALLATION OF DIGITAL UNIT
For ‘Digital Ready’ variants of the Tomey Slit Lamp a Digital Camera Assembly (DCA) must be
purchased separately, contact Tomey or your local distributor for details.
Set up the Slit Lamp as directed in the Instructions For Use (IFU) EP59-70030 or EP59-51500,
contact your distributor if you are unable to locate a copy of these instructions.
1. Install the DCA carefully remove the eyepiece assembly from the magnifi cation block by
unscrewing the securing knob at the same time as supporting the eyepieces. The
eyepieces assembly is a dovetail type fi t to the magnifi cation block.
Care must be taken to avoid getting any dirt or dust on any of the optical
components.
Securing knob
2. Fit the DCA to the rear of the magnifi cation block and tighten the securing knob to hold
it in place.
3. Re-fi t the eyepiece assembly to the rear of the DCA and secure it by tightening the
securing knob.
4. Rotate the swing arm to either side of the unit, and with a suitable Philips screwdriver
remove the cable channel cover plate.