ENVIRONMENTAL CONDITIONS
Ambient temperature from 5 to 40° C.
Relative humidity max. 80% for temperatures up to 31° C with linear decrease up to
50% at the temperature of 40° C condensing included.
Atmospheric Pressure from 500 to 1060 hPa.
Voltage supply variation not higher than
10%.
Value of transitory over-voltage in conformity to the installation category (which
provides for 2500V).
INSTALLATION
The device has been calibrated and tested at factory and as such does not require any
further calibration or adjustments before installation and start-up.
Unpack the device and follow the next advices:
1.
Position the device on a levelled flat, hard and smooth surface, established with non-
infiammable material.
2.
Allow space not lower than 10 cm from walls and surrounding furniture.
3.
Do not install the device nearby water sinks or similar to avoid contacts with water
or substances which could cause electrical short-circuit to the system.
4.
Install the device in a well aerated location, not near windows or external doors which
could cause an unnatural circulation of the air inside the device and therefore
compromise its correctly running.
5.
Do not install the device nearby heat sources or near other electrical devices.
6.
Install the device in such a way that the power cable is never twisted or bent, but it
should connect to the socket free and unhindered. Avoid to positioning the power cable
nearby heat sources or other devices that shall caused damage on it in a long run.
7.
In the event the device is placed on a trolley, always check that the lower part of the
device is not enclosed or hindered, as to allow continuos and sufficient ventilation.
Once the device is correctly installed and power cable connected, it’s ready for use.
SAFETY WARNINGS
The device is designed for use in internal locations.
The device is not designed for use in presence of gasses or explosive vapours.
No water or other liquids should be poured into device neither on its base.
Before any cleaning or maintenance actions, the power cable must always first be
removed.
Make sure that the electrical system has electrical grounding and that it’s in conformity
to the safety norms in the country of installation.
Do not remove any label or plate, in case of need, ask for more.
Always insist on original spare parts.
Do not open the door of the device until the thermometer of the internal temperature do
not indicate a temperature lower than 30°C.
Medical devices, before being placed in the sterilizer, should be washed and dried. The
residual water steam may create deposits on the probes and resistances by altering the
precision
.
TIME – PROGRAMMED RUNNING
1.
Insert the supplied connection socket of the mains cable in the plug of the device
(Pos. 2) and the feeder- plug in the wall mains socket, after checking the voltage value.
2.
Place materials to be sterilized inside the device
3.
Lock up the door. (Pos. 5)
4.
Keep the air-inlets half-opened (Pos. 3) to allow a better circulation of the air inside
the device and to distribute the heat even to the corners and hidden spaces. Close the
above air-inlets only at the end of the sterilization cycle in order to seal the device and
to keep the material inside in temperature for a long time.
5.
The device is switch on by pressing the button ON/OFF (Pos. 7).
6.
Once the device is on, the display TEMP. °C (Pos. 9) shows the current internal
temperature, while the display TIME (Pos. 10) shows the last set timing for
sterilization cycle.
7.
For programming or changing sterilization time, press the button SET (Pos. 8).
In this way the display TIME (Pos. 10) begin to blink. Use the buttons (Pos.
16) to change the value shown, up to sterilization time required. The length, expressed
in minutes, can be change from 1 to 250. At the end of blinking, the time shown shall
be automatically memorised (
ATTENTION! Let the blinking end by itself).
8.
To show the set heating temperature press the button SET (Pos. 8).
9.
To set or change temperature press SET (Pos. 8) for two consecutive times. In this
way the temperature display TEMP.°C (Pos. 9) will begin to blink. Use the
buttons (Pos. 16) to change the value shown, up to the temperature required.
The temperature can be changed from 1 to 200°C. At the end of blinking, the
temperature shown shall automatically be memorised and shall again be shown the
current internal temperature (
ATTENTION! Let the blinking end by itself).
(to be continued)