![titanox A3-216-400 Instruction Manual Download Page 5](http://html.mh-extra.com/html/titanox/a3-216-400/a3-216-400_instruction-manual_1120050005.webp)
TIME – PROGRAMMED RUNNING
(continued on)
10.
To activate the time-programmed sterilization running, press the button
START/STOP (Pos. 15). In this way the green LED (Pos. 14) will light up and
the set temperature shall appear on the display TEMP.°C (Pos. 9). If the button
START/STOP (Pos. 15) is kept pressed, the temperature shall remain visualised on
the display until the button is released. After releasing the button STAR/STOP (
Pos. 15), the green LED “resistances” (Pos. 13) will light up and display
TEMP.°C (Pos. 9) show the ward ECL (Electronic-Compensation-Low). This
indication shall remain until set temperature is reached.
11.
On reaching the set temperature the display TEMP.°C (Pos. 9) shall show the
ward ECH (Electronic-Compensation-High).
12.
At the end of the internal compensation phase, the display TEMP.°C (Pos. 9) shall
show the current internal temperature and shall begin the countdown, shown by the
blinking of the decimal point (last value on the right) on the display TIME (Pos.
10).
13.
Once sterilization cycle time is terminated, the green LED (Pos. 14) shuts itself
automatically.
*** ATTENTION ***
In the adjustment phase, the green LED (Pos. 13) indicates the activation of the
resistances for maintaining the temperature programmed. In case of abnormality of the
driving system of the resistances, on display TIME (Pos. 10), the indication FAL is
shown and the green LED (Pos. 13) shall remain off.
TECHNICAL SPECIFICATION
The device is in conformity to electrical safety norms provided for by the normative
institutes and supplied with bipolar plug which assures perfect electrical grounding.
NON-COMPLIANCE WITH INSTRUCTIONS DESCRIBED IN THIS BOOKLET SHALL
FREE COMPANY TITANOX S.R.L. FROM ANY LIABILITY.
SYMBOLS
MEANING
Warning !
See the annexed documentation
Caution! Hot surface (max. 210 °C)
It points the Manufacturer’s name
This symbol, together with the year, points the production date
RAEE symbol to handle electrical and electronical devices wastes
Available operation:
TIME-PROGRAMMED: timer programmable up to 4 hours
CONTINUOUS CYCLE: manual (without timer)
7
4
Models:
A3-216-400
A3-217-535
Loaded Chamber
3 kg
4 kg
External sizes:
Width mm
570
705
Height mm
400
475
Depth mm
345
450
Internal Sizes:
Width mm
405
535
Height mm
210
345
Depth mm
255
320
Weights:
Net weight kg.
13 kg
22 kg
Gross weight kg.
15 kg
25 kg
Electrical Characteristics:
Nominal tension (Voltage)
230 V
230 V
Nominal power (Watt)
450 W
950 W
Nominal frequency (Hz)
50/60 Hz
50/60 Hz
Net’s fuses (mm 5x20)
F5A-250 V
F5A-250 V