titanox A3-216-400 Instruction Manual Download Page 5

TIME – PROGRAMMED  RUNNING 

(continued on) 

 
10.

 

To  activate  the  time-programmed  sterilization  running,  press  the  button  

START/STOP       (Pos. 15). In this way the green LED     (Pos. 14) will light up and 
the  set  temperature  shall  appear  on  the  display  TEMP.°C      (Pos.  9).  If  the  button 
START/STOP       (Pos. 15) is kept pressed, the temperature shall remain visualised on 
the display until the button is released. After releasing the button STAR/STOP      (  
Pos.  15),  the  green  LED  “resistances”          (Pos.  13)  will  light  up  and  display        
TEMP.°C              (Pos.  9)  show  the  ward  ECL  (Electronic-Compensation-Low).  This 
indication shall remain until set temperature is reached. 

11.

 

On  reaching the set  temperature the display  TEMP.°C        (Pos. 9) shall show the 

ward ECH (Electronic-Compensation-High). 

12.

 

At the end of the internal compensation phase, the display TEMP.°C    (Pos. 9) shall 

show  the  current  internal  temperature  and  shall  begin  the  countdown,  shown  by  the 
blinking of the decimal point (last value on the right) on the display TIME      (Pos. 
10). 

13.

 

Once  sterilization  cycle  time  is  terminated,  the  green  LED      (Pos.  14)  shuts  itself 

automatically. 

 
 

 

*** ATTENTION *** 

 
In  the  adjustment  phase,  the  green  LED              (Pos.  13)  indicates  the  activation  of  the 
resistances  for  maintaining  the  temperature  programmed.  In  case  of  abnormality  of  the 
driving  system  of  the  resistances,  on  display  TIME            (Pos.  10),  the  indication  FAL  is 
shown and the green LED      (Pos. 13) shall remain off. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

TECHNICAL SPECIFICATION

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

The  device  is  in  conformity  to  electrical  safety  norms  provided  for  by  the  normative 
institutes and supplied with bipolar plug which assures perfect electrical grounding. 

 

NON-COMPLIANCE WITH INSTRUCTIONS DESCRIBED IN THIS BOOKLET SHALL 

FREE COMPANY TITANOX S.R.L. FROM  ANY LIABILITY.

  

 

SYMBOLS 

MEANING 

 

Warning !  

 

See the annexed documentation 

 

 

Caution! Hot surface (max. 210 °C) 

 

It points the Manufacturer’s  name 

 

This symbol, together with the year, points the production date 

 

RAEE symbol to handle electrical and electronical devices wastes  

Available operation: 

TIME-PROGRAMMED: timer programmable up to 4 hours 
CONTINUOUS CYCLE: manual (without timer) 

 

Models: 

A3-216-400 

A3-217-535 

Loaded Chamber 

3 kg 

4 kg 

External sizes: 

 

 

Width  mm 

570 

705 

Height  mm 

400 

475 

Depth  mm 

345 

450 

Internal  Sizes: 

 

 

Width  mm 

405 

535 

Height  mm 

210 

345 

Depth  mm 

255 

320 

Weights: 

 

 

Net  weight kg. 

13 kg 

22 kg 

Gross weight kg. 

15 kg 

25 kg 

Electrical  Characteristics: 

 

 

Nominal  tension (Voltage) 

230 V 

230 V 

Nominal  power  (Watt) 

450 W 

950 W 

Nominal  frequency  (Hz) 

50/60 Hz 

50/60 Hz 

Net’s  fuses (mm  5x20) 

F5A-250 V 

F5A-250 V 

 

Summary of Contents for A3-216-400

Page 1: ...ning For this period Titanox s r l guarantees the availability of spare parts and safe operation as long as the environmental and use conditions defined in the instructions for use are respected by the user The supplier of the electronic card states that the capacity of the memory of the mounted microprocessor is 1 000 000 cycles By cycle we mean every ignition of the card therefore in 10 years 36...

Page 2: ...working of the device 6 The warranty expires when the device is tampered with or repaired by unauthorised personnel 7 It is excluded the substitution and the extension of the warranty following a breakdown 8 It is excluded any compensation for damages direct or indirect of any nature to persons or objects arising from use or suspension of use of the device 9 The warranty expires immediately if the...

Page 3: ...rials whose melting point is higher than 300 C surgical instruments plates and metal screw Inside the sterilizer place only metal containers without plastic parts and without textile material FORBIDDEN USES It is forbidden to place in the device items whose melting temperature is lower than 300 C or is not known to the operator ACCESSORY PARTS 2 Internal shelves 1 Mains cable SELLING OFF The steri...

Page 4: ...on display TEMP C Pos 9 shall appear the ward ECH Electronic Compensation High 11 At the end of the internal compensation working on display TEMP C Pos 9 shall appear the current internal temperature that persist with no time limit 12 To terminate the sterilization procedure in continuous cycle running press the button SELECT Pos 12 PLEASE NOTE If change of heating procedure is required re check e...

Page 5: ...e activation of the resistances for maintaining the temperature programmed In case of abnormality of the driving system of the resistances on display TIME Pos 10 the indication FAL is shown and the green LED Pos 13 shall remain off TECHNICAL SPECIFICATION The device is in conformity to electrical safety norms provided for by the normative institutes and supplied with bipolar plug which assures per...

Page 6: ... system has electrical grounding and that it s in conformity to the safety norms in the country of installation Do not remove any label or plate in case of need ask for more Always insist on original spare parts Do not open the door of the device until the thermometer of the internal temperature do not indicate a temperature lower than 30 C Medical devices before being placed in the sterilizer sho...

Reviews: