ENVIRONMENTAL CONDITIONS 

 

Ambient temperature from 5 to 40° C. 

 

Relative humidity max. 80% for temperatures up to 31° C with linear decrease up to 
50% at the temperature of 40° C condensing included. 

 

Atmospheric Pressure from 500 to 1060 hPa. 

 

Voltage supply variation not higher than 

±

 10%.

 

 

Value  of  transitory  over-voltage  in  conformity  to  the  installation  category  (which 
provides for  2500V).

 

 

INSTALLATION 

The  device  has  been  calibrated  and  tested  at  factory  and  as  such  does  not  require  any 
further calibration or adjustments before installation and start-up. 
Unpack the device and follow the next advices: 
1.

 

Position the device on a levelled flat, hard and smooth surface, established with non-
infiammable material. 

2.

 

Allow space not lower than 10 cm from walls and surrounding furniture. 

3.

 

 

 

Do  not  install  the  device  nearby  water  sinks  or  similar  to  avoid  contacts  with 

water or substances which could cause electrical short-circuit to the system. 

4.

 

Install the device in a well aerated location, not near windows or external doors which 
could  cause  an  unnatural  circulation  of  the  air  inside  the  device  and  therefore 
compromise its correctly running. 

5.

 

 

 

Do not install the device nearby heat sources or near other electrical devices. 

6.

 

Install  the  device  in  such  a  way  that  the  power  cable  is  never  twisted  or  bent,  but  it 
should connect to the socket free and unhindered. Avoid to positioning the power cable 
nearby heat sources or other devices that shall caused damage on it in a long run. 

7.

 

In the event the device is placed on a trolley, always check that the lower part of the 
device is not enclosed or hindered, as to allow continuos and sufficient ventilation. 

Once the device is correctly installed and power cable connected, it’s ready for use.  

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

SAFETY WARNINGS 

 

The device is designed for use in internal locations.

 

 

The device is not designed for use in presence of gasses or explosive vapours.

 

 

No water or other liquids should be poured into device neither on its base. 

 

Before  any  cleaning  or  maintenance  actions,  the  power  cable  must  always  first  be 

removed. 

 

Make sure that the electrical system has electrical grounding and that it’s in conformity 

to the safety norms in  the country of installation. 

 

Do not remove any label or plate, in case of need, ask for more. 

 

Always insist on original spare parts. 

 

Do not open the door of the device until the thermometer of the internal temperature do 

not indicate a temperature lower than 30°C. 

 

TIME – PROGRAMMED  RUNNING 

1.

 

Insert  the  supplied  connection  socket  of  the  mains  cable  in  the  plug  of  the  device  
(Pos. 4) and the feeder- plug in the wall mains socket, after checking the voltage value. 

2.

 

Turn the thermoregulator knob (Pos. 1) and set the desired temperature. This operation  shall 
always be carried out when device is cold. 

3.

 

Place materials to be sterilized inside the device. 

4.

 

Lock up the door (Pos. 8). 

5.

 

Set,  by turning the timer knob (Pos. 2) anti-clockwise, the sterilization time adding necessary 
time  in  order  to  bring  the  device  and  the  material  inside  to  the  desired  temperature.

  The 

necessary time to reach the temperature of 180°C is about 30/40 minutes in case of empty 
chamber  sterilizer,  and  about  60  minutes  with  Loaded  Chamber  well  arranged  on  the 
trays.

 This operation switch on the warning light (Pos. 5) indicating  that the device is working 

and the warning light “resistance” (Pos. 6) indicating that the resistances are turned on  in the 
circuit.  The  graduated  thermometer  (Pos.  3)    allows  to  check  the  progressive  increase  of  the 
internal  temperature.  When  the  internal  temperature  has  reached  the  programmed  value,  the 
warning  light  “resistance”  (Pos.  6)  is  turned  off  for  then,  after  brief  time,  again  turned-on 
indicating  the  re-connection  of  the  resistances  in  the  circuit.  This  operation  allows  to  keep 
inside  the  device  a  constant  temperature  up  to  the  end  of  the  sterilization  cycle.  After 
programmed time has elapsed, the device turns itself off automatically.  
In  order  to  maintain  an  even  temperature  inside  of  the  heating chamber and to allow a better 
circulation of the air, the device is featured with a ventilation motor fan internally wired. 

 

CONTINUOS -CYCLE  RUNNING 

The device has also a continuos-cycle running. For use this kind of running proceed up to 
point 4 of the procedure for time-programmed running described above. Turn, at this 
moment, the timer knob (Pos. 2) anti-clockwise. The indefinite-time manual continuos-
cycle is inserted with a click. To turn off the device, turn back the timer know to position 0.

 

 

Summary of Contents for A3-213-400V

Page 1: ...C CR R I IT TA AL LY Y V Vi ia a C Ca an no ov ve e 2 2 A A C Ca an no ov ve e d dè è B Bi ia az zz zi i T Te el l 0 00 03 39 9 0 03 37 75 5 3 39 94 40 06 65 5 r r a a F Fa ax x 0 00 03 39 9 0 03 37 75 5 3 39 94 40 06 67 7 H Ho om me e p pa ag ge e h ht tt tp p w ww ww w t ti it ta an no ox x c co om m E E M Ma ai il l i in nf fo o t ti it ta an no ox x c co om m INSTRUCTIONS MANUAL DRY HEAT STERI...

Page 2: ...ndicated in these booklet instructions or in any case results from phenomenon unrelated to the normal working of the device 6 The warranty expires when the device is tampered with or repaired by unauthorised personnel 7 It is excluded the substitution and the extension of the warranty following a breakdown 8 It is excluded any compensation for damages direct or indirect of any nature to persons or...

Page 3: ...boratory tests following a cycle of 120 minutes at 180ºC on Bacillus subtilis var niger ATCC 9372 spores demonstrated the efficiency of the sterilizer Titanox does not guarantee and cannot ensure the effective achievement of the steril state of the instruments placed in the sterilizer according to the definition of steril medical instruments forseen by the EN 556 Norms The user of the sterilizer t...

Page 4: ...s in conformity to the safety norms in the country of installation Do not remove any label or plate in case of need ask for more Always insist on original spare parts Do not open the door of the device until the thermometer of the internal temperature do not indicate a temperature lower than 30 C TIME PROGRAMMED RUNNING 1 Insert the supplied connection socket of the mains cable in the plug of the ...

Reviews:

Related manuals for A3-213-400V