ORDINARY MAINTENANCE
Before initiating any maintenance operation, make sure that:
•
the device is not connected to the power source.
•
the device is at ambient temperature.
Keep the internal parts and the grid perfectly clean. Even though they may change colour and
become brownish, never use abrasive or inflammable products for cleaning them.
Keep the external parts perfectly clean that they have to be always specular to prevent corrosion and
dust.
PERIODIC MONTHLY MAINTENANCE
After removing the plug from the power socket, check that the fuses are not oxidised specially when
the device is not used for a long time or kept in a humid ambient.
The power socket should not change colour or oxidise. If that happens, replace it immediately.
The power cable must be integral and it should not show cuts, abrasions or bending.
The resistances and the internal electrical system do not require any maintenance.
PERMITTED USES
The device must be used to sterilize metal materials whose point of melting is higher than 300°C
(surgical instruments, plates, and metal screw).
FORBIDDEN USES
It is forbidden to place in the device items whose temperature of melting is lower than
300°C or it’s not known to the operator.
ACCESSORY PARTS
2 Internal shelves
1 Mains cable
SELLING OFF
The sterilizing unit is made of various materials with mechanical, electro-mechanical and electronic
parts.
The selling off has to be made according with the regulations in force in the utilizing Nation.
***ATTENTION***
Laboratory tests following a cycle of 120 minutes at 180ºC on Bacillus subtilis var niger ATCC
9372 spores demonstrated the efficiency of the sterilizer.
Titanox does not guarantee and cannot ensure the effective achievement of the steril state of the
instruments placed in the sterilizer, according to the definition of steril medical instruments forseen
by the EN 556 Norms. The user of the sterilizer therefore has the responsability to conduct all of the
confirmation procedures of the sterilization process and the necessary verifications in order to ensure
the effective completion of every single sterilization cycle.
TECHNICAL SPECIFICATIONS
The device is in conformity to electrical safety norms provided for by the normative
institutes and supplied with bipolar plug which assures perfect electrical grounding.
NON-COMPLIANCE WITH INSTRUCTIONS DESCRIBED IN THIS
BOOKLET SHALL FREE COMPANY TITANOX S.R.L. FROM ANY
LIABILITY.
Device type B.
Caution! See the annexed documentation.
Caution! Hot surface (max. 210°C).
Available operation:
TIME-PROGRAMMED: timer programmable up to 2 hours
CONTINUOUS CYCLE: manual (without timer)
5
2
Models:
A3-213-400V
A3-214-535V
A3-215-670V
Loaded Chamber
3 kg
4 kg
5 kg
External sizes:
Width mm
565
705
835
Height mm
410
475
550
Depth mm
315
420
555
Internal Sizes:
Width mm
405
535
670
Height mm
220
350
430
Depth mm
250
325
425
Weights:
Net weight kg.
13 kg
22 kg
35 kg
Gross weight kg.
15 kg
25 kg
41 kg
Electrical Characteristics:
Nominal tension (Voltage)
230 V
230 V
230 V
Nominal power (Watt)
450 W
950 W
1900 W
Nominal frequency (Hz)
50/60 Hz
50/60 Hz
50/60 Hz
Net’s fuses (mm 5x20)
F5A-250 V
F5A-250 V
F10A-250 V
Ventilation motor (fan)
23 W
23 W
23 W