Gas Entrainment or Modulating Fluids
Fluids which entrap gasses or seriously modulate the ultrasound will cause problems with the time of
flight measurement. This might include viscous fluids being re-circulated from a small volume
reservoir or Glycol type fluids.
Gasses seriously affect the ultrasound and should be avoided. It is recommended that the meter is
always installed in a positive pressure system with a back pressure of at least 0.5 Bar plus 2 times the
fluid vapour pressure. It is preferable to install the meter in a way that any trapped vapour can escape
the flow measurement section. i.e. in a vertical line with the flow upwards.
The meters are calibrated on water at ambient temperature typically 21 to 25°C.