The seventh flow in the table above is zero. This could have been a small bubble passing through the
flow meter bore which could completely absorb the ultrasound. The median filter completely ignores
this reading and for the zero reading returns the value of 3 from the results either side. This filter is
particularly useful at low flows where a small dip in value could drop a result below the internal cut off
levels.