INTRODUCTION & SAFETY  

 

TIS Checkbox is a professional, multifunctional field checkbox, intended for use with all makes and models of 
installation test instruments. 

In addition to its durable body and simple operation it boasts the following features:  

  Insulation calibration with test voltages up to 1000 V,  

 

LowΩ and continuity accuracy verification,  

  Continuity and insulation of measurement leads verification,  

 

Hi Current and “No-trip loop” impedance measurement verification  

  Verification of RCD trip-out time measurement, including test current verification - 30 mA X1/X5. 

  Voltage and frequency calibration (A calibrated Multimeter to be used to verify mains voltage & 

frequency). 
Recommended Multimeter(s):  
                           

TIS201                                                         TIS280 

 
 

 

 

 

 

 

 

 

                                

 

Please visit our web site: www.testinstrumentsolutions.co.uk to explore the full range of test instruments we 
provide.  

 

 

The TIS Checkbox is designed for accuracy verification whilst on-site testing, this product does not 
replace the need to have the instrument calibrated periodically.

  

 
(Please refer to the manufacturers recommended calibration interval) 

 

 

Summary of Contents for TIS CHECKBOX

Page 1: ...TIS CHECKBOX NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSON S INSTRUCTION MANUAL VERSION 1 00...

Page 2: ...ended Verification Sequence Record Form 11 Warranty Information 12 Optional Accessories Test Leads Part Number TIS172ML Description Universal three wire 4mm 13A Mains lead Part Number TIS173DB Descrip...

Page 3: ...loop impedance measurement verification Verification of RCD trip out time measurement including test current verification 30 mA X1 X5 Voltage and frequency calibration A calibrated Multimeter to be u...

Page 4: ...ckbox Damaged leads accessories must not be used in conjunction with the TIS Checkbox as this may result in a safety hazard damage to the TIS Checkbox Do not use the TIS Checkbox on supply systems wit...

Page 5: ...4 CHECKBOX DESCRIPTION MAINTENANCE Figure 1 0 Figure 1 1 1 2 3 8 4 5 6 9 10 11 12 7 13 14 15 16 17...

Page 6: ...st current is too low 7 RCD 30mA X5 test calibration function selected 8 RCD Fault loop selection button Press repeatedly to cycle through available options Local LOOP LOOP 0 5 LOOP 2 LOOP 50 RCD 30mA...

Page 7: ...ith soapy water or alcohol Then leave the instrument to dry totally before use Warning Do not use liquids based on flammable cleaning fluids or hydrocarbons Do not spill cleaning liquid over the instr...

Page 8: ...dance Verification Local Loop Connect Instrument under test to the 230V 13A socket Figure 1 0 Section13 Check polarity of mains input connect TIS Checkbox to mains outlet Select Local Loop range using...

Page 9: ...g the RCD LOOP button Fig 1 Section 8 Perform either Hi Current or NO TRIP Impedance test Check that the loop impedance measurement has increased from the local loop impedance reading by a nominal of...

Page 10: ...t conducting this test however if the RCD current fails please refer to manufacturers specifications RCD Test Verification 30 00mA X1 Select RCD 30 00mA X1 range using the RCD LOOP button Fig 1 Sectio...

Page 11: ...ds to Null Zero the lead resistance Com Null 0 500R 5 Maximum input current 350mA 5 000R 2 Maximum input current 350mA 10 000R 2 Maximum input current 350mA Insulation Resistance 1 000MR 1 Maximum inp...

Page 12: ...Insulation Resistance 250V 1 00M M 50 00M M Insulation Resistance 500V 1 00M M 50 00M M 200 0M M Insulation Resistance 1000V 1 00M M 50 00M M 200 0M M AC Voltage DMM Voltage Reading V Meter Under Test...

Page 13: ...purchase invoice may be requested to validate the purchase date Any unauthorized repair adjustment will void the warranty For service calibration repair requirements please contact UNIT 12 14 LUDDITE...

Reviews: