Tinsley 5867 Operating Instructions Manual Download Page 4

TYPE 5867
PORTABLE ‘Q’ METER

OPERATING INSTRUCTIONS

5867 Operating Instructions : Issue 3

www.tinsley.co.uk

4

5. MAINTENANCE

Normally no maintenance is required other than cleaning with a moist cloth. Avoid aggressive detergents or solvents.

CAUTION:

Before any maintenance, repair or exchange of parts or fuses, the instrument must be disconnected from 
the mains supply and all power sources. In the event of a fault occurring, the instrument should be returned 
to our factory or agent. A mains fuse is fitted to the mains inlet socket on the front panel, and should be 
replaced if necessary.

CAUTION:

Disconnect mains lead and all connecting leads, before removing fuse holder. Replace only with the correct 
fuse type, i.ee according to the following chart.

6. BATTERY CHARGING

The 5867 has built-in rechargeable sealed lead acid batteries which are fully charged when delivered. To indicate the
state of charge, 8 LED’s indicate the remaining battery capacity in % of full charge. 

The battery charger is built-in and the instrument may be connected to a voltage supply of between 90 to 260V AC. The
instrument may be operated with a flat battery, provided that the unit is powered to the mains – in this way the unit can be
simultaneously operated and charged. If the unit is switched on and the battery is not sufficiently charged, the unit will
turn itself off automatically. To conserve the battery, if no key is pressed during a 20 minute period the unit will also switch
itself off. To protect the battery from complete discharge below a minimum capacity the unit will self power off.

Charging is automatically controlled with built-in protection circuits eliminating the possibility of over charging. 

The unit will be fully recharged in 8 hours without being used at the same time.

A fully charged unit should allow continuous operation for a working day, without the need to connect to the mains.

BATTERIES

The internal batteries are a sealed lead acid type. Care should be taken when disposing of them and they may be
returned to Tinsley for safe disposal.

Line Voltage 

Range VAC 47-63Hz

Fuse (250V) IEC 127

Selection

5 x 20mm

100V

87-110V

1A (T)

120V

104-132V

220V

191-242V

1A (T)

240V

209-264V

Summary of Contents for 5867

Page 1: ...S TINSLEY A DIVISION OF HARTEST PRECISION INSTRUMENTS LTD SUBJECT TO CHANGE WITHOUT NOTICE This manual superseded all previous versions please keep for future reference IMPORTANT Please read these ins...

Page 2: ...features warning lights which are illuminated when appropriate Accessories supplied as standard 1 mains cord 1 operating instructions English 2 test leads colour coded length 3m each supplied with sel...

Page 3: ...The use of makeshift fuses and the short circuiting of fuse holders is prohibited Whenever it is likely that the protection has been impaired the apparatus shall be made inoperative and be secured aga...

Page 4: ...s built in and the instrument may be connected to a voltage supply of between 90 to 260V AC The instrument may be operated with a flat battery provided that the unit is powered to the mains in this wa...

Page 5: ...temperature of the unit The display will then show which capacitance range and measuring range was last used and this will be shown on the display until it is next changed by the operator The last use...

Page 6: ...dicating that a higher range should be selected which can be achieved by pressing the up arrow key A suitable range as would be chosen by the Auto range method described above would be to pick a range...

Page 7: ...titled Oven This will flash when the oven is within 5 C of the standard 40 C REMOTE This illuminates when the Q meter is being controlled remotely and is lit whilst commands are being transferred via...

Page 8: ...nds with xn T 84dec Get Internal Temperature TempCo xC C 67dec Do Calibration cycle A 65dec Get Internal Voltages I 73dec Get Unit Itent string Tinsley 5867 xVx L OPT Start logging readings QPT Output...

Page 9: ...ices at any time without prior notice For further details please refer to our website www tinsley co uk 1 Warner Drive Springwood Industrial Estate Braintree Essex CM7 2YW Tel 01376 335271 E mail info...

Reviews: