Calibration 

The model APS voltmeter is accurate to +/- 5% of the output voltage, as shown on the LCD 

display on the instrument panel. It is recommended that this model follow an annual calibration 

cycle to ensure the instrument is in good working order and that the LCD of the integration 

voltmeter is accurate. 

 

 

Battery Charging 

The Model APS comes with two batteries. The lithium-ion battery should be charged after each 

use. The Model APS will indicate when the battery voltage is low, by a slow blinking of the 

Green Power LED, instead of providing a steady glow. The battery charger has an LED. RED color 

LED indicates the battery is being charged. GREEN color LED indicates that battery is fully 

charged. GREEN color LED also means that the battery is on trickle charge, and can be left in 

this condition. The charger can be used with 110v / 240v AC. It is an auto sensing charger that 

can be used throughout the world. Please consult your manual if you have a battery with a 
serial number prior to 827. 

 

 

Summary of Contents for APS

Page 1: ...c model supplies 800V to 35kV in one instrument with two ranges of 3 500V to 35 000V and a low range of 800V to 8 000V It is recommended for use on any pipe diameter as well as on flat surfaces Tinker Rasor Model APS Unpacking and Set up Inspection should be made upon receipt to ensure that all parts were received unscathed and every piece of your equipment is in working order First remove all acc...

Page 2: ...c These types of electrodes are ideal for inspecting pipeline coatings in a plant or the field These are square cut and tightly wound for tight fit and intimate contact to coatings and will inspect the entire coated surface of the pipe in one pass Silicone rubber electrodes are made of conductive material with very low resistivity a unique combination of silicone rubber and carbon It meets many di...

Page 3: ...ork fine in most cases unless the environment is very dry Place the live electrode on the test surface and move it over the work area at a constant speed as defined by the test method or standard you are working to Always be sure to switch off the instrument before disconnecting any cables The speed of the electrode s travel along the pipe should never be excessive since faulty inspection may resu...

Page 4: ...of the holiday detector This display gives the user the ability to tune the APS to a specific voltage with in increment in 100V steps Correct voltage output for a given thickness of coating has long been a matter of controversy However recent formulas have been suggested which may be used as a guide for correct peak voltages on various coating thickness The calculation is as follows NACE Internati...

Page 5: ...e charged after each use The Model APS will indicate when the battery voltage is low by a slow blinking of the Green Power LED instead of providing a steady glow The battery charger has an LED RED color LED indicates the battery is being charged GREEN color LED indicates that battery is fully charged GREEN color LED also means that the battery is on trickle charge and can be left in this condition...

Reviews: